Writer Roller/Carriage Assembly; Removal; Reassembly - GE MAC 5000 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Writer Roller/Carriage Assembly

Removal

Reassembly

Revision B
Maintenance: Disassembly Guidelines
7. Connect cable to the main PCB.
8. Replace the display/keyboard assembly and the top cover following
procedures above.
1. Remove the power supply assembly following procedures above.
2. Inside the power supply compartment, disconnect the cable that
connects to the writer assembly.
3. Open the unit to access the paper compartment. Move the paper size
bracket to the A4 position to expose one of the writer assembly
mounting screws.
4. Remove the screw and return the paper size bracket to the
8.5 x 11 position.
5. Close the unit and turn it over so the bottom side is up.
6. Remove the four screws located on the underside of the writer roller/
carriage assembly and lift the writer from the bottom of the unit.
Reassemble the writer roller/carriage assembly by reversing the above
procedures.
MAC™ 5000 resting ECG analysis system
2024917-010
4-21