Status LED indications: (Continued)
LED mode and color
Steady green
Flashing green
Flashing red
Steady red

Enabling the leak sensors

When you power-on the system, the leak sensors default to disabled
Rule:
status unless previously enabled.
To enable the leak sensors:
1.
In the ACQUITY UPLC Console, select Control > Leak Sensors.
2.
To enable the leak sensor for an individual module, click the status on
the left-hand side of the module's description.
To enable all leak sensors, click Enable All.
Tip:
3-4
Preparing the System
Description
• Binary solvent manager – Indicates solvent is
flowing.
• Sample manager – Indicates the sample
manager is operating normally, attempting to
complete any outstanding samples or
diagnostic function requests. When sample
and diagnostic function requests are finished,
the LED reverts to the unlit mode.
• Detector – Indicates the detector lamp is
ignited.
• Binary solvent manager and sample manager
– Indicates that the module is initializing.
• Detector – Indicates the detector is initializing
or calibrating.
Indicates that an error stopped the instrument or
device. Refer to the ACQUITY UPLC Console for
information regarding the error.
Indicates a module's failure that prevents its
further operation. Power-off the module, and
then power-on. If the LED is still steady red,
contact your Waters service representative.