Figure 25. Details on the Instrument settings tab.
Note: The display brightness should not be switched off completely. This will change the screen to
black which can only be reverted by a system restart.
4.3.6

Performing camera exposure calibration

Note: Only Administrators can perform the camera exposure calibration.
WARNING
Moving parts
To avoid contact with moving parts during the operation of the EZ2, the
instrument must be operated with the hood closed.
If the hood sensor or lock is not functioning correctly, contact QIAGEN Technical
Services.
1. Before first use in the final installation site, the camera exposure calibration must be
performed.
2. From the Maintenance menu under the Camera LED tab, follow the instructions on the user
interface.
EZ2 Connect MDx User Manual 05/2022
54