WARNING
Toxic fumes
Do not use bleach to clean or disinfect the EZ2 instrument or used labware.
Bleach in contact with salts from the buffers can produce toxic fumes.
CAUTION
Damage to the instrument
Do not use spray bottles containing alcohol or disinfectant to clean surfaces of
the EZ2. Spray bottles should be used only to clean items that have been
removed from the worktable and if permitted by local laboratory operating
practices.
WARNING
Risk of fire
Do not allow cleaning fluid or decontamination agents to come into contact
with the electrical parts of the EZ2.
WARNING
Risk of electric shock
Do not open any panels on the EZ2.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
WARNING
Risk of fire or explosion
When using ethanol or ethanol-based liquids to clean the EZ2, handle such
liquids carefully and in accordance with the required safety regulations. If any
liquid spills, wipe it off and leave the EZ2 hood open to allow flammable
vapors to disperse.
EZ2 Connect MDx User Manual 05/2022