9
Technical Specifications
QIAGEN reserves the right to change specifications at any time.
9.1
Operating conditions
Power
Fuse
Overvoltage category
Air temperature
Relative humidity
Altitude
Place of operation
Pollution level
Environmental class
Average noise level (over 8 hours)
IP Code (IEC 60529)
9.2
Transport conditions
Air temperature
Relative humidity
Environmental class
9.3
Storage conditions
Air temperature
Relative humidity
Environmental class
EZ2 Connect MDx User Manual 05/2022
100−240 V AC, 50/60 Hz, 1000 VA
Mains supply voltage fluctuations are not to exceed 10%
of nominal supply voltages.
AC Inlet: T4A H 250 V
Heater block (temperature fuse): 10A 250V 117C
II
18–30°C (64–86°F)
10–75% RH
Up to 2000 m (6500 ft.)
For indoor use only
2
3K21 (IEC 60721-3-3)
3M11 (IEC 60721-3-3)
Max. 70 dBA
IP20
−25−60ºC (−13−140ºF) in manufacturer's package
Note: If the EZ2 is transported in temperatures below 0ºC
(32ºF), it is recommended to wait 24 hours before
switching on the instrument to allow it to reach the
temperature conditions of the installation environment.
5–85% RH
2K11 (IEC 60721-3-2)
2M4 (IEC 60721-3-2)
5−40°C (41−104°F) in manufacturer's package
5−85% RH
2K11 (IEC 60721-3-2)
2M4 (IEC 60721-3-2)
164