3. Put on gloves
4. Clean the piercing unit (if this is not yet done during after run maintenance).
o
To prepare the piercing unit for cleaning, tap Move down.
o
Open the hood.
o
Clean the piercing unit as recommended for After run maintenance (see page 145).
5. Clean the worktable with 70% ethanol, and then with distilled water (if this is not yet done
during after run maintenance).
6. Close the hood.
7. To return the piercing unit to home position, tap Move up.
8. To allow tray removal, tap Move to back.
9. Clean the tray with 70% ethanol, and then with distilled water.
Note: After removal of the tray for cleaning purposes, ensure it is correctly reinstalled.
Figure 149-A Right side tray installed.