Power Requirements - Qiagen EZ2 Connect MDx User Manual

Hide thumbs Also See for EZ2 Connect MDx:
Table of Contents

Advertisement

CAUTION
Risk of overheating
To ensure proper ventilation, maintain a minimum clearance of 10 cm at the
sides and rear of the EZ2.
Slits and openings that ensure the ventilation of the instrument must not be
covered.
WARNING
Risk of personal injury and material damage
The EZ2 is too heavy to be lifted by one person. To avoid personal injury or
damage to the instrument, do not lift the instrument alone.
CAUTION
Damage to the instrument
Direct sunlight may bleach parts of the instrument, cause damage to plastic
parts, and may interfere with the load check.
The EZ2 must be located out of direct sunlight.
CAUTION
Damage to the instrument
Do not use the EZ2 in the vicinity of sources of strong electromagnetic radiation
(for example unshielded, deliberately operated high-frequency sources or
mobile radio devices), because these can interfere with the proper operation.
4.1.2

Power requirements

The EZ2 operates at: 100–240 V AC +/– 10%, 50/60 Hz, 1000 VA
Make sure the voltage rating of the EZ2 is compatible with the AC voltage available at the
installation site.
EZ2 Connect MDx User Manual 05/2022
45