6.6

Camera calibration

The EZ2 Connect MDx provides an internal camera which allows an inventory scan prior to run
execution.
Note: The camera calibration should be performed as part of installation procedure, after moving
the instrument, as well as in case of load check issues.
1. On the touchscreen, tap Maintenance.
2. Tap Camera LED.
Figure 155. Camera exposure calibration screen.
3. Follow the instructions displayed on the screen.
4. Close the hood.
5. To begin the procedure, tap Start.
EZ2 Connect MDx User Manual 05/2022