5.8

Aborting a protocol run

A protocol run can be stopped at any time. To stop the run, follow the steps below:
1. In the Protocol run in progress screen, tap Abort run. A confirmation dialog appears.
2. Tap Yes to stop the run, or tap No to go back to the Protocol run in progress screen.
Figure 128. Aborting protocol run dialog.
3. When the run is aborted, the instrument finishes the current movement, and then attempts to
dispense the contents of the pipettes into the first available empty tubes, and to release tips
into empty tip holders. Then, the worktable returns to its initial position. After this is done, a
message is shown and the Proceed to the summary button is activated. Tap Proceed to the
summary.
Note: If you tap Abort during a pause or when the machine is waiting to reach a specific
temperature, the run is stopped immediately.
Figure 129. Protocol run interrupted dialog.
4. Tap Finish to end the run and go back to the Home screen. A run report is generated. For
more information on how to save a run report, refer to the Saving a run report section (see
page 128).