WARNING
Risk of electric shock
Do not open any panels on the EZ2.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
Any other maintenance or repair may only be carried out by authorized field
service.
To ensure satisfactory and safe operation of the EZ2, follow the advice below:
The line power cord must be connected to a line power outlet that has a protective conductor
(earth/ground).
Place instrument in a location so that the power cord is accessible and can be readily
connected/disconnected.
Use only the power cord delivered by QIAGEN.
Do not adjust or replace internal parts of the instrument.
Do not operate the instrument with any covers or parts removed.
If liquid has spilled inside the instrument, switch off the instrument, and if the spill has not been
completely contained in the tray, disconnect it from the power outlet, and contact QIAGEN
Technical Services for advice before attempting any troubleshooting.
If the instrument becomes electrically unsafe, prevent other personnel from operating it and contact
QIAGEN Technical Services.
The instrument may be electrically unsafe when:
The EZ2 or the line power cord appear to be damaged.
The EZ2 has been stored in unfavorable conditions for a prolonged period.
The EZ2 has been subjected to severe transport stresses.
Liquids have come into direct contact with electrical components of the EZ2.
The power cord has been exchanged with a non-official power cord.
WARNING
Electric hazard
Do not touch the EZ2 with wet hands.