Qiagen EZ2 Connect MDx User Manual page 153

Hide thumbs Also See for EZ2 Connect MDx:
Table of Contents

Advertisement

Figure 150. Weekly maintenance screen.
3. Put on gloves.
4. Clean the O-rings with a lint free tissue to remove any previously applied grease.
5. Using a fresh glove, apply a small amount of silicon grease onto the surface of the O-rings
(shown in Figure 150), by just using a fingertip moistened with grease. Care should be taken
not to allow grease to be applied to, or cover the opening in the nub of the pipettor.
6. Apply a small amount of silicon grease (see Appendix B – EZ2 Connect MDx Accessories,
Ordering Information, page 170) onto the inner wall of the large end of a fresh filter-tip by
using the fine end of a second fresh tip.
Figure 151. Preparation of filter tips for grease application.
EZ2 Connect MDx User Manual 05/2022
153

Advertisement

Table of Contents
loading

This manual is also suitable for:

9003230

Table of Contents