Calibration; Preparation; Enter Calibration Mode; Select Tube Type And Units - Teledyne HPM 4 Instruction Manual

Table of Contents

Advertisement

Digital AVC Instruction Manual

Calibration

The HPM 4/5/6 is calibrated for all three tube types (DV-4, DV-5, and DV-6) at the factory. This ensures valid
readings will be displayed for each tube type when delivered. The user may re-calibrate the HPM 4/5/6 if
desired using either a reference tube or a vacuum gauge tube at a known vacuum using the instructions below.
3.5.1.

Preparation

Ensure the HPM 4/5/6 is turned off.
Plug in a DV-4, DV-5 or DV-6 Reference tube or plug into a tube at a known vacuum within the valid
range of the tube (see 1.2 for tube ranges).
3.5.2.

Enter Calibration Mode

Push and hold "TUBE SELECT" and "UNITS SELECT" simultaneously.
While holding these buttons, press and hold the "Power" button.
When the LCD displays "CALIBRATION MODE", release all three buttons.
The LCD will display "CAL MODE" at the top and will flash "Select Tube Type" while flashing the tube
type in the upper right-hand corner.
3.5.3.

Select Tube Type and Units

Press "TUBE SELECT" or "UNITS SELECT" until the correct gauge tube is displayed.
Press "POWER" to accept the tube type selection.
The LCD will flash "Select Units" while flashing the Pressure Units in the lower right-hand corner
Press "TUBE SELECT" or "UNITS SELECT" until the desired pressure units are displayed.
Press "POWER" to accept the unit selection.
The LCD will now display the current pressure in the selected units.
3.5.4.

Adjustment

The user may now adjust the displayed pressure value by pressing "TUBE SELECT" (Up) to increase or
"UNITS SELECT" (Down) to decrease.
Use the buttons to match the displayed pressure to the known vacuum level.
While adjusting, a message may be displayed informing the user that they must wait before further
adjustment is allowed.
Allow time for the reading to settle (DV-5 tubes take about 3 minutes).
When the desired pressure value has been reached, press the "POWER" button to set this selection
and exit calibration mode.
NOTE: The table in 1.3.2 specifies which Teledyne Hastings reference tube to be used in calibration.
EAR99 Technology Subject to Restrictions Contained on the Cover Page