When a component's pre-set interval expires, the Maintenance LED on the scanner's control
panel will light, the tool icon will change from blue to yellow, and that component's bar graph will
be red. In addition, an alert for that specific component will display in the form of a triangular
yellow exclamation mark icon (see image - specifically "Clean Transport" or either "Front Camera"
or "Back Camera").
N
OTE
The presence of a red bar graph denotes that maintenance is required for that specific
component based on the pre-set interval. The presence of a red bar graph will not disrupt or
physically stop the scanner's normal function or operation.
Resetting the Maintenance Reminders
If a component's maintenance interval expires triggering the maintenance reminders, a hyperlink
for that component will navigate the Operator to the reset process required to restart the interval
count. This reset process is necessary to ensure required maintenance continues to be
monitored appropriately (instructions pertaining to the function of each component's hyperlink are
detailed fully in
"Maintenance Procedures" on page
The maintenance reminders can be singularly disabled if the Operator so chooses; however BBH
does not recommend doing so.
N
OTE
The alerts for calibrating the front and back cameras are hard coded in the scanner and not
adjustable by the Operator. If an alert to calibrate the camera(s) is displayed, click the
corresponding hyperlink to reset the reminder and begin calibration. For instructions on how
to calibrate the cameras, go to
36
59).
"5.3.4 Camera Calibration" on page 80
in this manual.