KSK 150 FC
Safety check
CAUTION
R082
Version 7.1
G-30-1158-en
Risk of electric shock!
If cables and plug connections come loose, housing parts can come into con-
tact with live voltage. If the user or the patient touches these live housing
parts, they can receive an electric shock.
• Make sure that the regular technical safety checks required for this sys-
tem in accordance with the applicable national regulations are performed
on schedule and to the stipulated extent.
The safety checks may only be carried out by the manufacturer or by qualified
persons.
The scope of the safety checks of the system should comprise at least the fol-
lowing points:
–
Availability of the Instructions for Use
–
Visual inspection of the system and accessories for damage and legibility
of the labels
–
Leakage current test
–
Test of protective ground conductor
–
Function and wear test of the steerable casters and locking tabs
–
Function test of all switches, buttons, sockets and indicator lamps of the
system
Care and Maintenance
Page 61