Causes Of Malfunctions And Remedies - Zeiss KSK 150 FC Instructions For Use Manual

Table of Contents

Advertisement

What to do in the event of malfunctions
What to do in the event of malfunctions

Causes of malfunctions and remedies

Problem
No electrical function at all (il-
lumination does not work).
Illumination on colposcope
not working.
Page 56
For your safety
• This system is a high-grade technological product. To ensure optimum
performance and safe working order, we recommend having it checked
by our service representative as part of regular scheduled maintenance.
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions", attach a sign to the system
stating it is out of order and contact our service representative.
Possible cause
Power cord not plugged in.
Line failure.
Line voltage does not comply
with set voltage.
Defective device fuse.
Defective device cable.
Power switch not pressed.
Knob for brightness control
set to the left-hand stop, in
the "min" position.
Defective halogen lamp.
Corroded lamp contacts.
Remedy
• Connect the device cable. Page 46
• Contact in-house techni-
cian.
• Check the set power sup-
ply voltage.
• Change the fuse.
• Replace device cable.
• Press power switch.
• Turn the knob clockwise
towards "max".
• Change the halogen
lamp.
• Change the halogen
lamp.
KSK 150 FC
See
Page 46
Page 64
Page 28
Page 28
Page 62
Page 62