Zeiss OPMI Vario S88 Instructions For Use Manual

Zeiss OPMI Vario S88 Instructions For Use Manual

Hide thumbs Also See for OPMI Vario S88:
Table of Contents

Advertisement

S88 / OPMI
S8 / OPMI
S81 / OPMI
Instructions for Use
G-30-1607-en
Version 6.2
1/18/2013
Carl Zeiss
®
Vario
®
Vario
®
Vario

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OPMI Vario S88 and is the answer not in the manual?

Questions and answers

Summary of Contents for Zeiss OPMI Vario S88

  • Page 1 Carl Zeiss ® S88 / OPMI Vario ® S8 / OPMI Vario ® S81 / OPMI Vario Instructions for Use G-30-1607-en Version 6.2 1/18/2013...
  • Page 2 – S88 floor stand with lifting column, reference number: 000000-1169-820 – S81 ceiling mount, reference number: 000000-1176-969 – S8 ceiling mount, reference number: 000000-1176-968 ® – OPMI is a registered trademark of Carl Zeiss. Trademarks Manufacturer Carl Zeiss Meditec AG Goeschwitzer Strasse 51-52...
  • Page 3 S88, S8, S81 / OPMI Vario Chapter overview Chapter: Safety Measures Chapter: System Overview Chapter: Preparations for use Chapter: Operation Chapter: What to do in the event of malfunctions Chapter: Care and Maintenance Chapter: System Data Chapter: Index Version 6.2 G-30-1607-en Page 3...
  • Page 4 S88, S8, S81 / OPMI Vario Version 6.2 Page 4 G-30-1607-en...
  • Page 5: Table Of Contents

    S88, S8, S81 / OPMI Vario Safety Measures Safety Measures Key to symbols .................. 7 Hazard symbols ...................7 Information symbols..................7 Target group ..................8 Field of application ................8 Intended use ....................8 Normal use....................8 Liability and warranty................9 Notes for the operator..............10 Requirements for operation.............
  • Page 6 Safety Measures S88, S8, S81 / OPMI Vario Labels on the light sources of the suspension systems....... 54 Version 6.2 Page 6 G-30-1607-en...
  • Page 7: Key To Symbols

    S88, S8, S81 / OPMI Vario Safety Measures Key to symbols We would like to inform you about safety aspects which must be observed when handling this device. This chapter contains a summary of the most im- portant information concerning matters relevant to instrument safety. Hazard symbols The following safety information has been incorporated into the Instructions for Use.
  • Page 8: Target Group

    It is the duty of the customer or institution operating the system to train and instruct all staff using the system. Additional service activities are not part of these Instructions for Use. They will be performed by staff especially trained for this purpose by Carl Zeiss. Field of application Intended use...
  • Page 9: Liability And Warranty

    S88, S8, S81 / OPMI Vario Safety Measures CAUTION Not for diagnostic purposes! Do not use the recorded video sequences, video clips (cut sequences) and single images for diagnostic purposes. The visualized images may include de- viations in shape, contrast and color. Liability and warranty Warranty and liability depend on the applicable contractual stipulations.
  • Page 10: Notes For The Operator

    Safety Measures S88, S8, S81 / OPMI Vario Notes for the operator The correct use of the device is absolutely vital for safe operation. Therefore, please thoroughly familiarize yourself with the content of these Instructions for Use before starting up the device. Please also observe the Instructions for Use of additional instrument equipment.
  • Page 11 (see section "Care and Maintenance"). • Modifications and repairs of this device or any equipment operated to- gether with this device may only be performed by Carl Zeiss service staff or other persons authorized by Carl Zeiss.
  • Page 12 Safety Measures S88, S8, S81 / OPMI Vario • Avoid looking directly into the light source, e.g. into the microscope ob- jective lens or into the light guide, in order to prevent damage to the eye! • Start with the lowest brightness setting and gradually increase brightness up to the necessary and still admissible level.
  • Page 13: Requirements For Operation

    Safety Measures Requirements for operation Prior to the very first use Carl Zeiss Service or a person authorized by Carl Zeiss will install the device. Please make sure that the following requirements continue to be met for fur- ther operation: ...
  • Page 14: During Use

    Safety Measures S88, S8, S81 / OPMI Vario • Check the surgical microscope for sufficient freedom of movement. The device itself or accessories may be damaged if conflicting with each other. • To prevent accessories from falling down, check before every use that ac- cessories are securely locked in position and securing screws are firmly tightened.
  • Page 15: After Every Use

    S88, S8, S81 / OPMI Vario Safety Measures • Defective or unidentified accessories may lead to increased leakage cur- rent on the system and injure the patient. Never connect any defective or unidentified accessories and never touch the power outlet or video inter- faces while in contact with the patient.
  • Page 16: Connecting Navigation Systems (Option)

    – Implementing a procedure that guarantees the calibration of the sur- gical microscope which is absolutely vital for the use of the Carl Zeiss components "surgical microscope on suspension system" in combina- tion with a connected navigation system.
  • Page 17: Risk Of Burn Injuries Caused By High Illumination Intensity

    S88, S8, S81 / OPMI Vario Safety Measures • To check the accuracy of the overall system, perform the test specified by the navigation system manufacturer, e.g. the landmark test, also using the surgical microscope. This allows you to ensure that the stereotactic data has been correctly generated and transmitted to the navigation system without errors.
  • Page 18 – Certain medications affect the sensitivity to light. Recommendations Due to the large number of different factors involved and the lack of scientific publications on this topic, Carl Zeiss cannot provide guidance on acceptable illumination intensities and exposure durations. However, the surgical micro- Version 6.2...
  • Page 19 The area of the incision should be constantly irrigated. Final remark Carl Zeiss recommends to reduce illumination of the surgical field in that pos- sible extent to ensure the patient's security and a good microscopic image. Please refer to the warnings and precautions in this case.
  • Page 20: Safety Devices Of The Suspension Systems

    Safety Measures S88, S8, S81 / OPMI Vario Safety devices of the suspension systems 1 Release bar allows non-sterile persons to unlock the magnetic brakes of the suspen- sion system. 2 Adjustment screw for limiting downward travel serves to set the minimum working distance from the surgical field in ver- tical direction.
  • Page 21 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 21...
  • Page 22 Safety Measures S88, S8, S81 / OPMI Vario Stand lifting column 1 Key to set the optimum viewing height of the surgical microscope or for downward movement into the transport position. As long as you keep switch (1) pressed down, the lifting column (2) in the stand base moves upwards or downwards, depending on the position of the switch.
  • Page 23 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 23...
  • Page 24 Safety Measures S88, S8, S81 / OPMI Vario Xenon illumination system CAUTION The xenon lamp has a limited service life of 500 h! If used beyond its maximum service life, the xenon lamp may explode. • Please replace the xenon lamp in due time. •...
  • Page 25 S88, S8, S81 / OPMI Vario Safety Measures 4 Manual mode When the manual mode has been activated, all electrical control systems are disabled. The lamp brightness is automatically adjusted to a fixed set- ting. Version 6.2 G-30-1607-en Page 25...
  • Page 26 Safety Measures S88, S8, S81 / OPMI Vario Manual function 1 Manual button The Manual button permits you to switch to manual operation. The mo- torized functions of the surgical microscope are deactivated. The lamp brightness is automatically adjusted to a default setting. This lamp bright- ness value is displayed in the first display field..
  • Page 27 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 27...
  • Page 28: Symbols And Labels On The Device

    Safety Measures S88, S8, S81 / OPMI Vario Symbols and labels on the device CAUTION Note the warning labels and notes! • If you notice that any label is missing on your system or has become illeg- ible, please contact us or one of our authorized representatives. We will supply a replacement.
  • Page 29 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 29...
  • Page 30 Safety Measures S88, S8, S81 / OPMI Vario 5 Friction adjustment of vertical axis 6 Scale on focusing knob 7 Scale on illuminated-field knob 8 Device name Version 6.2 Page 30 G-30-1607-en...
  • Page 31 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 31...
  • Page 32 Safety Measures S88, S8, S81 / OPMI Vario 9 Zoom release button 10 Focus release button 11 Release button for magnetic brakes Version 6.2 Page 32 G-30-1607-en...
  • Page 33 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 33...
  • Page 34: Labels On The S88 Floor Stand

    Safety Measures S88, S8, S81 / OPMI Vario Labels on the S88 floor stand 1 "Maximum load" warning label The maximum load (accessory equipment) on the microscope body must not exceed 20 kg! 2 "Risk of crushing" warning label Fingers may be crushed. Do not touch this area while moving the surgical microscope or parts of it.
  • Page 35 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 35...
  • Page 36 Manufacturer (company name) – Your options to contact the device manufacturer, i.e., phone number, fax number and email address of the local contact of the national Carl Zeiss sales organization. XXXX XXXXXX – SIP No. A unique identification number assigned to your device.
  • Page 37 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 37...
  • Page 38 Safety Measures S88, S8, S81 / OPMI Vario 13 Height adjustment (optional) Identifies the direction of movement of the stand with lifting column for the respective switch position. 14 Lifting column activation time This label represents the maximum permissible operating time as well as the subsequent mandatory rest phase for the lifting column and indicates the ratio between the operating time and the rest phase.
  • Page 39 S88, S8, S81 / OPMI Vario Safety Measures 13, 14 Version 6.2 G-30-1607-en Page 39...
  • Page 40: Labels On S8 Ceiling Mount

    Safety Measures S88, S8, S81 / OPMI Vario Labels on S8 ceiling mount 1 "Maximum load" warning label The maximum load (accessory equipment) on the microscope body must not exceed 20 kg! 2 "Risk of crushing" warning label Fingers may be crushed. Do not touch this area while moving the surgical microscope or parts of it.
  • Page 41 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 41...
  • Page 42 Manufacturer (company name) – Your options to contact the device manufacturer, i.e., phone number, fax number and email address of the local contact of the national Carl Zeiss sales organization. XXXX XXXXXX – SIP No. A unique identification number assigned to your device.
  • Page 43 S88, S8, S81 / OPMI Vario Safety Measures 8, 9, 10 Version 6.2 G-30-1607-en Page 43...
  • Page 44 "Maintenance and care". 12 "Vertical movement" indicating label This label identifies the operating handle for the vertical movement of the ceiling mount. Contact Carl Zeiss Service when moving the ceiling mount to its standby position requires excessive force. Version 6.2 Page 44...
  • Page 45 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 45...
  • Page 46: Labels On S81 Ceiling Mount

    Safety Measures S88, S8, S81 / OPMI Vario Labels on S81 ceiling mount 1 "Maximum load" warning label The maximum load (accessory equipment) on the microscope body must not exceed 20 kg! 2 "Risk of crushing" warning label Fingers may be crushed. Do not touch this area while moving the surgical microscope or parts of it.
  • Page 47 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 47...
  • Page 48 Manufacturer (company name) – Your options to contact the device manufacturer, i.e., phone number, fax number and email address of the local contact of the national Carl Zeiss sales organization. XXXX XXXXXX – SIP No. A unique identification number assigned to your device.
  • Page 49 S88, S8, S81 / OPMI Vario Safety Measures 8, 9, 10 11 Version 6.2 G-30-1607-en Page 49...
  • Page 50: Labels On The Suspension System's Connector Panel

    Safety Measures S88, S8, S81 / OPMI Vario Labels on the suspension system's connector panel 1 "Remote connector" label Identifies a connector to which devices with a maximum breaking capacity of 24 V / 0.5 A may be connected. 2 "Connector of wired foot control panel" label 3 "Power switch S2"...
  • Page 51 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 51...
  • Page 52: Additional Label On The S88 Floor Stand (Instrument Tray Option)

    Safety Measures S88, S8, S81 / OPMI Vario Additional label on the S88 floor stand (Instrument tray option) 1 "Maximum load on instrument tray" indicating label The maximum load of accessories on the instrument tray must not exceed 13 kg. Version 6.2 Page 52 G-30-1607-en...
  • Page 53 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 53...
  • Page 54 Safety Measures S88, S8, S81 / OPMI Vario Labels on the light sources of the suspension systems 1 "Warning of a hazard area" indicating label The Superlux light source is equipped with an integrated xenon illumina- tion system. The xenon illumination must not be used for ophthalmic sur- gery.
  • Page 55 S88, S8, S81 / OPMI Vario Safety Measures Version 6.2 G-30-1607-en Page 55...
  • Page 56 Safety Measures S88, S8, S81 / OPMI Vario Version 6.2 Page 56 G-30-1607-en...
  • Page 57 S88, S8, S81 / OPMI Vario System Overview System Overview S88 / OPMI Vario ................60 Design .......................60 S88 / OPMI Vario with lifting column ..........62 Design .......................62 S8 / OPMI Vario ................64 Design .......................64 S81 / OPMI Vario ................66 Design .......................66 Surgical microscope.................
  • Page 58 System Overview S88, S8, S81 / OPMI Vario Video monitor (option) ................104 S88 floor stand with lifting column ..........112 Features....................112 Description of the modules ..............113 Design ....................114 Stand base with lifting column..............116 Connector panel ..................118 S8 ceiling mount ................120 Features....................
  • Page 59 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 59...
  • Page 60: S88 / Opmi Vario

    System Overview S88, S8, S81 / OPMI Vario S88 / OPMI Vario Design 1 Surgical microscope (see Page 69) 2 Coupling 3 S88 floor stand (see Page 94) Version 6.2 Page 60 G-30-1607-en...
  • Page 61 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 61...
  • Page 62: S88 / Opmi Vario With Lifting Column

    System Overview S88, S8, S81 / OPMI Vario S88 / OPMI Vario with lifting column Design 1 Surgical microscope (see Page 69) 2 Coupling 3 S88 floor stand with lifting column (see Page 112) Version 6.2 Page 62 G-30-1607-en...
  • Page 63 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 63...
  • Page 64: S8 / Opmi Vario

    System Overview S88, S8, S81 / OPMI Vario S8 / OPMI Vario Design 1 Surgical microscope (see Page 69) 2 Coupling 3 S8 ceiling mount (see Page 120) Version 6.2 Page 64 G-30-1607-en...
  • Page 65 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 65...
  • Page 66: S81 / Opmi Vario

    System Overview S88, S8, S81 / OPMI Vario S81 / OPMI Vario Design 1 Surgical microscope (see Page 69) 2 Coupling 3 S81 ceiling mount (see Page 126) Version 6.2 Page 66 G-30-1607-en...
  • Page 67 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 67...
  • Page 68 System Overview S88, S8, S81 / OPMI Vario Version 6.2 Page 68 G-30-1607-en...
  • Page 69: Surgical Microscope

    S88, S8, S81 / OPMI Vario System Overview Surgical microscope Special features The apochromatic optics of the OPMI Vario surgical microscope provide su- perb optical quality. The microscope image displays optimum contrast and su- perb detail recognition along with excellent depth of field. An integrated, mo- torized Varioskop objective lens allows the working distance from the surgical field to be adjusted between 200 and 415 mm.
  • Page 70: Design

    System Overview S88, S8, S81 / OPMI Vario Design The surgical microscope comprises the following modules: 1 Coupling for mounting the surgical microscope on the suspension system. 2 Carrier arm for the surgical microscope 3 Balancing system including magnetic brakes This system allows balancing of the surgical microscope.
  • Page 71 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 71...
  • Page 72: Controls, Displays, Connections

    System Overview S88, S8, S81 / OPMI Vario Controls, displays, connections 1 Friction adjustment of vertical axis Use this knob to adjust the friction of the vertical axis as required. 2 Balance setting of lateral tilt motion Use this knob to adjust the balance setting of the lateral tilt motion. 3 Balance setting of front-to-back tilt motion Use this knob to adjust the balance setting of the front-to-back tilt motion.
  • Page 73 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 73...
  • Page 74 System Overview S88, S8, S81 / OPMI Vario 6 Focus stop button This button permits you to deactivate the electrical drive of the focusing system. After you have pressed the focus stop button, you can only focus manually on the surgical field using knob (9). The focus stop button is lit. To release the stop, press the focus stop button again (light in the button goes out).
  • Page 75 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 75...
  • Page 76 System Overview S88, S8, S81 / OPMI Vario 11 Freely programmable release buttons Specific functions of the suspension system can be assigned to these but- tons, e.g.: increasing / reducing brightness etc. 12 Zoom release button for setting the magnification factor from 0.4x to 2.4x. 13 Focus release button for continuous focusing within the working distance of 200 to 415 mm.
  • Page 77 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 77...
  • Page 78 System Overview S88, S8, S81 / OPMI Vario X-Y coupling (option) The OPMI Vario can be equipped (and also retrofitted) with an X-Y coupling (2). Our service team or an authorized person will install the X-Y coupling for you. The X-Y coupling allows motorized fine positioning of the surgical microscope in a horizontal plane.
  • Page 79 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 79...
  • Page 80: Binocular Tubes And Eyepieces

    System Overview S88, S8, S81 / OPMI Vario Binocular tubes and eyepieces 180° tiltable tube 1 PD adjustment knob The correct position has been reached when the two eyepiece images merge into one. You can read off the interpupillary distance set on the ad- justment knob.
  • Page 81 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 81...
  • Page 82 System Overview S88, S8, S81 / OPMI Vario Widefield eyepieces with magnetic coupling NOTE Eyepieces have a magnetic field! Please note that the usual rules for the handling of magnets must be ob- served for eyepieces removed from the tube: •...
  • Page 83 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 83...
  • Page 84: Xenon Illumination System

    System Overview S88, S8, S81 / OPMI Vario Xenon illumination system CAUTION Injury to the patient's eye! Never use xenon illumination for ophthalmic procedures! • Make sure that no xenon light enters the patient's eyes. The suspension system is equipped with a xenon illumination system for fiber illumination.
  • Page 85 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 85...
  • Page 86 System Overview S88, S8, S81 / OPMI Vario When inserting a new lamp module, make sure that the knob (2) is set to “1". If the first lamp fails, switch to the second lamp in logical sequence. 3 Indicator: Backup lamp is in use When the segment in knob (2) lights up, the backup lamp is in use.
  • Page 87 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 87...
  • Page 88 System Overview S88, S8, S81 / OPMI Vario 8 Brightness control You can adjust the brightness using the two control keys on the control panel. The brightness of the xenon lamp can also be adjusted by pressing the ap- propriate buttons on the foot control panel. 9 Yellow indicator lamp is lit when the lamp has failed, or if the lamp module is defective.
  • Page 89 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 89...
  • Page 90: Identical Modules Of The Suspension Systems

    System Overview S88, S8, S81 / OPMI Vario Identical modules of the suspension systems Suspension arm 1 Lock of the cable duct – For opening, turn a quarter turn clockwise or counterclockwise. – For closing, press down and turn a quarter turn clockwise or counter- clockwise.
  • Page 91 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 91...
  • Page 92: Display Field With Control Keys

    System Overview S88, S8, S81 / OPMI Vario Display field with control keys The display and control panel is integrated in the control unit. The surgical microscope on the suspension system can be controlled either manually or electronically. The control software required for electronic control is installed in the electronics box of the suspension system.
  • Page 93 S88, S8, S81 / OPMI Vario System Overview Yellow LED above the "MANUAL" key The yellow LED is lit when you have switched to the manual mode. Version 6.2 G-30-1607-en Page 93...
  • Page 94: S88 Floor Stand

    System Overview S88, S8, S81 / OPMI Vario S88 floor stand Features The S88 floor stand is a carrier system for the surgical microscope. It is used to power and control the motorized functions of the surgical microscope. The hallmarks of the floor stand are its superb mobility and easy operation. Four steerable casters on the stand base permit easy positioning in the OR.
  • Page 95: Description Of The Modules

    S88, S8, S81 / OPMI Vario System Overview Description of the modules The S88 floor stand comprises the articulated arm, the stand column and the stand base. The articulated arm comprises a carrier arm and a suspension arm. The carrier arm contains the control unit with all electrical supply systems re- quired for the control of a motorized surgical microscope.
  • Page 96: Design

    System Overview S88, S8, S81 / OPMI Vario Design 1 Control panel 2 Carrier arm 3 Xenon illumination system, see Page 84) 4 Suspension arm 5 Stand base 6 Stand column Version 6.2 Page 96 G-30-1607-en...
  • Page 97 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 97...
  • Page 98: Stand Base With Column

    System Overview S88, S8, S81 / OPMI Vario Stand base with column 1 Maneuvering handle for moving the stand. 2 Support for hanging up the foot control panel during transport. 3 Cable support (2x) for winding up the power cord and the cable of the foot control panel. 4 Cable deflector protects the cables lying on the floor from damage caused by rolling over them with the stand base.
  • Page 99 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 99...
  • Page 100: Connector Panel

    CAUTION Danger! Electrical voltage! • Only connect accessories and medical devices designated by Carl Zeiss for use with this system to the power outlet sockets. When connecting other devices, make sure that safety is guaranteed regarding admissible touch currents and ground leakage currents as per IEC 60601-1.
  • Page 101 S88, S8, S81 / OPMI Vario System Overview 9 Strain relief device The strain relief device prevents inadvertent unplugging of the following electrical connections. – Power cable. – connecting cable for foot control panel or operating chair with appro- priate foot switch. Version 6.2 G-30-1607-en Page 101...
  • Page 102: Instrument Tray (Option)

    The instrument tray (1) can carry a maximum of 13 kg. The tray has been de- signed, for example, for mounting MediLive Trio from Zeiss: – The MediLive Trio is attached to the instrument tray with the aid of two stud bolts.
  • Page 103 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 103...
  • Page 104: Video Monitor (Option)

    • Do not use video images for diagnostic purposes as the video camera is not calibrated for these purposes. The background illumination of the LCD display has a limited service life. If you notice that the display is getting darker or starts to flicker, contact your Zeiss dealer. Version 6.2...
  • Page 105 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 105...
  • Page 106 System Overview S88, S8, S81 / OPMI Vario Components The principal component of the TFT monitor is the 15" screen which delivers flawless, sharp images even at low frame rates of 50 Hz. The connectors and controls are located under cable cover (2) on the back of the TFT monitor.
  • Page 107 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 107...
  • Page 108 System Overview S88, S8, S81 / OPMI Vario Connector panel CAUTION Risk of injury caused by wrong accessories! • Only operate the device with the power cable included in the delivery package! 2 Power supply for power and voltage supply of the device. 3 DVI connection DVI stands for Digital Video Interface and is the latest technology for dig- ital data transmission.
  • Page 109 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 109...
  • Page 110 System Overview S88, S8, S81 / OPMI Vario Powering on the TFT monitor To facilitate the operation of the TFT monitor, it is automatically activated when the suspension system is switched on. While the system is booted up, the TFT monitor executes a power-on se- quence that verifies the signals of all ports (DVI, VGA, S-Video, Composite and Component).
  • Page 111 S88, S8, S81 / OPMI Vario System Overview 360° 90° ±90° 45° Version 6.2 G-30-1607-en Page 111...
  • Page 112: S88 Floor Stand With Lifting Column

    System Overview S88, S8, S81 / OPMI Vario S88 floor stand with lifting column Features The S88 floor stand with lifting column is a suspension system for surgical mi- croscopes. It is used to power and control the motorized functions of the sur- gical microscope.
  • Page 113: Description Of The Modules

    S88, S8, S81 / OPMI Vario System Overview Description of the modules The S88 floor stand comprises of the articulated arm, the motorized stand lifting column and the stand base. The articulated arm comprises a carrier arm and a suspension arm. The carrier arm contains the control panel with all electrical supply systems re- quired for the control of a motorized surgical microscope.
  • Page 114: Design

    System Overview S88, S8, S81 / OPMI Vario Design 1 Control unit 2 Carrier arm 3 Lamp housing with Xenon illumination 4 Suspension arm 5 Motorized stand lifting column 6 Stand base Version 6.2 Page 114 G-30-1607-en...
  • Page 115 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 115...
  • Page 116: Stand Base With Lifting Column

    System Overview S88, S8, S81 / OPMI Vario Stand base with lifting column 1 Lifting column 2 Handgrip for moving the floor stand. 3 Support for hanging up the foot control panel during transport. 4 Up/down switch for setting the optimum viewing height of the surgical microscope, for moving the system up into the standby position or down into the trans- port position.
  • Page 117 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 117...
  • Page 118: Connector Panel

    CAUTION Danger! Electrical voltage! • Only connect accessories and medical devices designated by Carl Zeiss for use with this system to the power outlet sockets. When connecting other devices, make sure that safety is guaranteed regarding admissible touch currents and ground leakage currents as per IEC 60601-1.
  • Page 119 S88, S8, S81 / OPMI Vario System Overview 9 Strain relief device The strain relief device prevents inadvertent unplugging of the following electrical connections. – Power cable. – connecting cable for foot control panel or operating chair with appro- priate foot switch. Version 6.2 G-30-1607-en Page 119...
  • Page 120: S8 Ceiling Mount

    System Overview S88, S8, S81 / OPMI Vario S8 ceiling mount Features The S8 ceiling mount is a carrier system for the surgical microscope. It is used to power and control the motorized functions of the surgical microscope. The hallmarks of the S8 ceiling mount are its superb mobility and easy operation. The motorized functions can be activated using a foot control panel.
  • Page 121: Description Of The Modules

    S88, S8, S81 / OPMI Vario System Overview Description of the modules The S8 ceiling mount comprises an articulated arm, a suspension arm with the illumination system and a control unit. The articulated arm consists of a lift arm and a carrier arm. The lifting function permits the ceiling mount to be moved to a standby position.
  • Page 122: Design

    System Overview S88, S8, S81 / OPMI Vario Design 1 Lift arm 2 Carrier arm 3 Suspension arm 4 Handgrip for moving the ceiling mount into the standby or working position. 5 Xenon illumination system (see Page 84) 6 Control panel Version 6.2 Page 122 G-30-1607-en...
  • Page 123 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 123...
  • Page 124: Power Switch With Connector (Option)

    System Overview S88, S8, S81 / OPMI Vario Power switch with connector (option) The power switch and the connector can be either installed in the OR, or they can be integrated in the ceiling mount, at the back of the carrier arm (see il- lustration).
  • Page 125 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 125...
  • Page 126: Ceiling Mount S81

    System Overview S88, S8, S81 / OPMI Vario Ceiling mount S81 Features The S81 ceiling mount is a carrier system for the surgical microscope. It is used to power and control the motorized functions of the surgical microscope. The hallmarks of the S81 ceiling mount are its superb mobility and easy operation. The motorized functions can be activated using a foot control panel.
  • Page 127: Description Of The Modules

    S88, S8, S81 / OPMI Vario System Overview Description of the modules The S81 ceiling mount comprises a column, a carrier arm and a suspension arm. The suspension arm with the illumination system and the control unit is mounted on the carrier arm. The control unit is rotatable through 180° and contains all electrical supply systems required for the control of a motorized surgical microscope.
  • Page 128: Design

    System Overview S88, S8, S81 / OPMI Vario Design 1 Column 2 Carrier arm 3 Suspension arm 4 Xenon illumination system (see Page 84) 5 Control panel Version 6.2 Page 128 G-30-1607-en...
  • Page 129 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 129...
  • Page 130: Power Switch, Connector And Socket (Option)

    System Overview S88, S8, S81 / OPMI Vario Power switch, connector and socket (option) The connector and socket can be either installed in the OR, or they can be in- tegrated in the ceiling mount, at the back of the carrier arm (see illustration). 1 Rail The delivery package contains a cable clip which is used to guide the cable of the foot control panel away from the operating table.
  • Page 131 S88, S8, S81 / OPMI Vario System Overview Version 6.2 G-30-1607-en Page 131...
  • Page 132 System Overview S88, S8, S81 / OPMI Vario Version 6.2 Page 132 G-30-1607-en...
  • Page 133 S88, S8, S81 / OPMI Vario Preparations for use Preparations for use Attaching the equipment...............134 Mounting the surgical microscope............134 Attaching accessories ................136 Mounting the tube and eyepieces............138 Connections................... 140 Connecting the surgical microscope ............140 Mounting the light guide.................140 Aligning the X-Y coupling................142 Strain relief device on S88 floor stand............144 Connecting the S88 floor stand ...............146 Relocating the unit ................
  • Page 134: Attaching The Equipment

    Preparations for use S88, S8, S81 / OPMI Vario Attaching the equipment CAUTION Risk of injury caused by lowering of the surgical microscope or accessories falling down! • Never change modules and accessories during a surgical procedure or above the patient. •...
  • Page 135 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 135...
  • Page 136: Attaching Accessories

    Preparations for use S88, S8, S81 / OPMI Vario Attaching accessories The comprehensive accessory program permits the OPMI Vario surgical micro- scope to be configured to your specific needs. Make sure not to exceed the maximum admissible load in order to ensure op- timum mobility and reliable operation of your OPMI Vario.
  • Page 137 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 137...
  • Page 138: Mounting The Tube And Eyepieces

    Preparations for use S88, S8, S81 / OPMI Vario Mounting the tube and eyepieces • Loosen securing screw (4) by a few turns. • Remove cover (3) and store it in a safe place. • Place binocular tube (2) on the surgical microscope and firmly tighten se- curing screw (4).
  • Page 139 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 139...
  • Page 140: Connections

    Preparations for use S88, S8, S81 / OPMI Vario Connections Connecting the surgical microscope • Turn locking cap (1) by a quarter turn to the right or left and pull up cover (2). • Plug microscope connector (3) into socket (4) on the suspension system and firmly tighten the securing screws on the connector.
  • Page 141 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 141...
  • Page 142: Aligning The X-Y Coupling

    Preparations for use S88, S8, S81 / OPMI Vario Aligning the X-Y coupling The OPMI Vario can be equipped (and also retrofitted) with an X-Y coupling. Our service team or an authorized person will install the X-Y coupling for you. The X-Y coupling allows motorized fine positioning of the surgical microscope in a horizontal plane.
  • Page 143 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 143...
  • Page 144: Strain Relief Device On S88 Floor Stand

    Preparations for use S88, S8, S81 / OPMI Vario Strain relief device on S88 floor stand You can secure the power plug and the multipoint connector of the switching component against inadvertent loosening by installing the two cables in strain relief device (1).
  • Page 145 S88, S8, S81 / OPMI Vario Preparations for use 320 mm Version 6.2 G-30-1607-en Page 145...
  • Page 146: Connecting The S88 Floor Stand

    Preparations for use S88, S8, S81 / OPMI Vario Connecting the S88 floor stand • Check the voltage indicated at (3). The suspension system is set at the factory to the rated voltage used in the country of destination. The rated voltage indicated at window (3) must correspond to the rated voltage available on the site of installation.
  • Page 147 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 147...
  • Page 148: Relocating The Unit

    Preparations for use S88, S8, S81 / OPMI Vario Relocating the unit CAUTION Risk of crushing - mind your fingers! Fingers may be crushed in the areas marked with the "risk of crushing" label. • Do not touch these areas while the system is being moved or brought in its working / transport position.
  • Page 149 Transport over long distances! Over long distances (e.g. removal, return for repair, etc.) the device must al- ways be transported in the original packaging or in special return packaging. • For details, please contact your dealer or the Carl Zeiss Service. Version 6.2...
  • Page 150: Adjusting The Supension System

    Preparations for use S88, S8, S81 / OPMI Vario Adjusting the supension system S88 floor stand with lifting column - Setting the ergonomic working height Bring the surgical microscope into a position convenient for you before each surgical procedure, and set the optimum ergonomic working height of the suspension system via the lifting column (without patient!).
  • Page 151 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 151...
  • Page 152: Adjusting The Balance Setting Of The Suspension Arm

    Preparations for use S88, S8, S81 / OPMI Vario Adjusting the balance setting of the suspension arm CAUTION Risk of crushing - mind your fingers! Fingers may be crushed in the areas marked with the "risk of crushing" label. • Do not touch these areas while the system is being moved or brought in its working / transport position.
  • Page 153 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 153...
  • Page 154: Adjusting The Limit Of Downward Travel

    Preparations for use S88, S8, S81 / OPMI Vario Adjusting the limit of downward travel CAUTION Risk of injury caused by lowering of the surgical microscope! • Do not adjust the position of the suspension arm over the patient. • Always limit the downward travel of the suspension arm to prevent any contact with the patient in case the surgical microscope is lowered acci- dentally.
  • Page 155 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 155...
  • Page 156: Positioning The S8 Ceiling Mount

    • Move the system into the standby position after use and make sure that sufficient headroom is available. • Contact Carl Zeiss Service when moving the suspension system to its standby position requires some force. Version 6.2 Page 156...
  • Page 157 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 157...
  • Page 158: Attaching Sterile Drapes

    Preparations for use S88, S8, S81 / OPMI Vario Attaching sterile drapes CAUTION The use of third party drapes may affect navigation systems Certain third party drapes (e.g. drapes with curved protective lens covers) may influence the accuracy of navigation systems due to an optical distortion caused by the convex shaped protective lens cover.
  • Page 159: Settings On The Control And Display Panel

    S88, S8, S81 / OPMI Vario Preparations for use Settings on the control and display panel Setting up the suspension system • Turn on the suspension system at its power switch. • Successively select the following functions on the suspension system: –...
  • Page 160: Balancing The Surgical Microscope

    Preparations for use S88, S8, S81 / OPMI Vario Balancing the surgical microscope NOTE Risk of injury caused by lowering of the surgical microscope! • Balance the completely equipped surgical microscope before surgery without the patient. • Hold the surgical microscope tightly at its handgrips before releasing the magnetic brakes.
  • Page 161 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 161...
  • Page 162: Adjusting The Surgical Microscope

    Preparations for use S88, S8, S81 / OPMI Vario Adjusting the surgical microscope Notes on adjustment – It is recommended to provide the interpupillary distance as well as the re- fraction values for the different users to adjust the surgical microscope more quickly and to preset the values in the scope of the preparation.
  • Page 163 S88, S8, S81 / OPMI Vario Preparations for use  Now, the microscope is adjusted in such a way that the image is sharply defined over the complete magnification range without any renewed fo- cusing during the magnification setting. If this is not the case, repeat the procedure. •...
  • Page 164 Preparations for use S88, S8, S81 / OPMI Vario Version 6.2 Page 164 G-30-1607-en...
  • Page 165: Surgical Microscope With A Laser Micromanipulator

    S88, S8, S81 / OPMI Vario Preparations for use Surgical microscope with a laser micromanipulator Please observe the separate Instructions for Use of your laser micromanipu- lator. NOTE Risk of injury caused by laser micromanipulator! It is absolutely vital that the focal planes of the surgical microscope and of the laser micromanipulator are identical.
  • Page 166 Preparations for use S88, S8, S81 / OPMI Vario Adjusting the surgical microscope and laser micromanipulator to the same focal plane The OPMI Vario is equipped with a motorized Varioskop zoom system which is operated via the focusing buttons of handgrip (1) or the foot control panel. The Varioskop is used for the motorized setting of the working distance (coarse focus) and the motorized adjustment of image definition (fine focus).
  • Page 167 S88, S8, S81 / OPMI Vario Preparations for use Version 6.2 G-30-1607-en Page 167...
  • Page 168 Preparations for use S88, S8, S81 / OPMI Vario Version 6.2 Page 168 G-30-1607-en...
  • Page 169 S88, S8, S81 / OPMI Vario Operation Operation Checklist ..................170 Positioning the S88 floor stand ............. 176 Using the display and key field ............178 General functions ..................178 OPMI Vario on the suspension system, user interface with SpeedFokus option ................182 Procedure ..................
  • Page 170: Checklist

    If a function fails, do not use the system for safety reasons. • If possible, correct the malfunction (see “What to do in the event of mal- functions“) or contact the Zeiss service department. Always check the function of the system before surgery (without patient!)
  • Page 171 S88, S8, S81 / OPMI Vario Operation Version 6.2 G-30-1607-en Page 171...
  • Page 172 Operation S88, S8, S81 / OPMI Vario Zoom  Check this function using the appropriate key. Focus  Check this function using the appropriate key. Knobs  Check the friction of the knobs for zoom (5), focusing (6) and illuminated- field size (7).
  • Page 173 S88, S8, S81 / OPMI Vario Operation Version 6.2 G-30-1607-en Page 173...
  • Page 174 Operation S88, S8, S81 / OPMI Vario Suspension systems After the system has been switched on, it automatically performs a self-test. The self-test is completed after approx. five seconds. xenon illumination  A beep sound is emitted after activation. The sound stops on error-free ig- nition of the xenon lamp.
  • Page 175 S88, S8, S81 / OPMI Vario Operation S88 floor stand only: Lifting column  Check that the upward/downward movement function is OK.  Check that the surgical microscope is in a convenient position and that you have set the optimum ergonomic working height of the stand via the lifting column.
  • Page 176: Positioning The S88 Floor Stand

    Operation S88, S8, S81 / OPMI Vario Positioning the S88 floor stand Please also read the chapter: "Relocating the device" (see Page 148). • Unlock the activated locking tabs (2). NOTE Unintentional movement of the system! • Position the suspension system in such a way that you can slide it away from the patient at any time.
  • Page 177 S88, S8, S81 / OPMI Vario Operation Version 6.2 G-30-1607-en Page 177...
  • Page 178: Using The Display And Key Field

    Operation S88, S8, S81 / OPMI Vario Using the display and key field General functions Operating keys "∇" and "Δ" The keys "∇" and "Δ" are always assigned to the display field on their left. Every time you press the "∇" key, the displayed value is decremented in pre- defined steps down to a certain minimum.
  • Page 179 S88, S8, S81 / OPMI Vario Operation The surgical microscope can no longer be operated via the foot control panel, the handgrips or the display and key field. In the manual mode, you can only switch the illumination on and off on the foot control panel and release the magnetic brakes by pressing the appro- priate keys on the surgical microscope.
  • Page 180 Operation S88, S8, S81 / OPMI Vario Basic mode Basic mode The basic mode is always displayed in the normal operating status and after power-on of the system. In the basic mode, the following settings are displayed, depending on the in- stalled surgical microscope: In the upper display field the current lamp brightness...
  • Page 181 S88, S8, S81 / OPMI Vario Operation Acoustic signals Three successive – Error message during the software check after power- beeps on of the suspension system. – Error message for an internal system error. One beep When the focus or zoom position is saved. One beep When brightness level 1.0 is reached.
  • Page 182: Opmi Vario On The Suspension System, User Interface With Speedfokus Option

    Operation S88, S8, S81 / OPMI Vario OPMI Vario on the suspension system, user interface with SpeedFokus option Comments on the menu overview: Basic mode Jump to next mode Speed mode Return to basic mode Autofocus If no button is pressed for 20 seconds, the MENU A mode FOCUS...
  • Page 183 S88, S8, S81 / OPMI Vario Operation Overview: Control functions for OPMI Vario with SpeedFokus (autofocus). Basic mode Speed mode Autofocus mode see Page 180 (Menu A), see Page 187 (Menu B), see Page 188 – Setting the lamp brightness –...
  • Page 184 Operation S88, S8, S81 / OPMI Vario – Configuration menus 4 and 5 (SETTINGS 4 and 5), complete "STORE" key The focusing and zoom values currently set on the surgical microscope are saved for the user currently selected. The function of the "STORE" key is the same in all modes.
  • Page 185 S88, S8, S81 / OPMI Vario Operation Basic mode for OPMI Vario Basic mode The basic mode for OPMI Vario comprises the following functions: – Setting the lamp brightness – Switching the display between working distance and total magnification – Setting the user ID, menu item "USER".
  • Page 186 Operation S88, S8, S81 / OPMI Vario Display of the working distance or total magnification This function permits you to switch between displays. Basic mode In the basic mode, either the working distance (in mm) or the total magnifica- tion (as a factor) is displayed in the middle display field. The associated buttons "−...
  • Page 187 S88, S8, S81 / OPMI Vario Operation Speed mode (MENU A) This mode permits you to set the adjustment speeds for the following micro- Speed mode scope functions: – Focusing Menu A – Zoom function FOCUS SPEED: – Adjustment of the X-Y coupling ZOOM SPEED: Path: Every time you switch on the system, the basic mode is automatically...
  • Page 188 Operation S88, S8, S81 / OPMI Vario SpeedFokus autofocus option The "SpeedFokus (Autofocus)" option enables you to select an object in the ROI (Region Of Interest, measuring field) in the image of the surgical field and to automatically focus it. Prerequisite for the SpeedFokus (autofocus) option: a suitable video system must be installed on your surgical microscope and must be switched on.
  • Page 189 S88, S8, S81 / OPMI Vario Operation If you have selected "Brake AF-auto, OFF", focusing by the SpeedFokus (auto- Autofocus mode focus) option will not start automatically when the magnetic brakes are locked, i.e. when you let go of the release buttons You can trigger the focusing process using the handgrips or the foot control Menu B Brake...
  • Page 190 Operation S88, S8, S81 / OPMI Vario Configuration mode 1 (SETTINGS 1) Configuration Configuration mode 1 permits you: mode 1 – to assign functions to buttons C and D of the foot control panel, SETTINGS 1 menu items "FOOT LEFT C:" and "FOOT RIGHT D:" FOOT LEFT C: –...
  • Page 191 S88, S8, S81 / OPMI Vario Operation Roll-over procedure: → AF → XY-RES → → FOC-MEM ↑ ↓ ← AUX ← PHOTO ← ZOOM-MEM ← If you have assigned FOC-MEM to key C or D of the foot control panel, you can determine by the length of time for which you press the relevant key of the foot control panel during operation whether a positioning run is to be trig- gered (press the key for less than 2 seconds) or whether the current position...
  • Page 192 Operation S88, S8, S81 / OPMI Vario If you press button (2) for less than 2 seconds, zoom positioning is triggered; if you press the button for more than 2 seconds, the current zoom position is saved. Path: To return to the basic mode, press the MODE button in configuration mode 1 (SETTINGS 1), or do not press any button for 20 seconds.
  • Page 193 S88, S8, S81 / OPMI Vario Operation Configuration mode 2 (SETTINGS 2) Configuration mode 2 permits you: Configuration mode 2 – to set the adjustment speeds of focus and X-Y coupling as a function of the zoom setting, menu item "SPEED FUNCT:", –...
  • Page 194 Operation S88, S8, S81 / OPMI Vario The dynamic speed control is based on the basic speed selected for focusing and the X-Y coupling in the speed mode. If you have already set a high adjustment speed for focusing and the X-Y cou- pling in the speed mode, dynamic speed control may possibly not be effective across the entire zoom range, as the maximum adjustment speed is reached beforehand.
  • Page 195 S88, S8, S81 / OPMI Vario Operation If you have selected a relatively high basic brightness in the basic mode, the automatic function may not be able to sufficiently compensate for reduced transmission of the optical system. Defining the power-on behavior of the system The "POW ON MODE"...
  • Page 196 Operation S88, S8, S81 / OPMI Vario Configuration mode 3 (SETTINGS 3) Configuration mode 3 permits you: Configuration mode 3 – to enter the eyepiece magnification for the calculation of the total magni- fication, menu item "EYE PIECE:" – to select the magnetic brakes to be unlocked, menu item "SELECT SETTINGS 3 EYE- BRAKES:"...
  • Page 197 S88, S8, S81 / OPMI Vario Operation Selecting the magnetic brakes to be unlocked You can define which magnetic brakes should be unlocked when you press button B (3) of the left and/or right handgrip. Menu item "SELECT BRAKES:" offers you the following options for selection: XY - Z The magnetic brakes for X-Y movement and Z movement can be unlocked separately by pressing button B (3) on the left or right...
  • Page 198 Operation S88, S8, S81 / OPMI Vario Path: To return to the basic mode, press the MODE button in the configura- tion mode, or do not press any button for 20 seconds. To change to configuration mode 4 (SETTINGS 4), simultaneously press the MODE and STORE buttons in configuration mode 3.
  • Page 199 S88, S8, S81 / OPMI Vario Operation Configuration mode 4 (SETTINGS 4) Configuration mode 4 Configuration mode 4 permits you: – to define the horizontal and vertical ROI position, SETTINGS 4 menu items "AF-ROI X-POS." and "AF-ROI Y-POS." AF-ROI X-POS. –...
  • Page 200 Operation S88, S8, S81 / OPMI Vario Path: To return to the basic mode, press the MODE button in the configura- tion mode, or do not press any button for 20 seconds. To change to configuration mode 5 (SETTINGS 5), simultaneously press the MODE and STORE buttons in configuration mode 4.
  • Page 201 S88, S8, S81 / OPMI Vario Operation Configuration mode 5 (SETTINGS 5) Menu item "AF-CCD SIZE" in configuration mode 5 permits you to enter the size of the CCD array. Path: Every time you switch on the system, the basic mode is automatically displayed.
  • Page 202: Procedure

    Operation S88, S8, S81 / OPMI Vario Procedure • Turn on the system at the power switch of the suspension system. • Set the required brightness of the surgical field illumination on the suspen- sion system. • Check the system using the checklist. •...
  • Page 203 S88, S8, S81 / OPMI Vario What to do in the event of malfunctions What to do in the event of malfunctions Failure of main functions ...............204 Failure of lamp control system ..............204 Failure of the focusing function ...............204 Failure of a xenon lamp ................205 Failure of magnetic brakes...............208 Failure of the zoom function..............209 Failure of the focusing function ...............210...
  • Page 204: Failure Of Main Functions

    What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Failure of main functions NOTE Failure of main functions! In the event that one of the major functions fails (XY movement, focus, zoom, light control), leading to the impairment of further functions, you can switch to the manual mode to complete an already started surgical proce- dure.
  • Page 205: Failure Of A Xenon Lamp

    S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Failure of a xenon lamp CAUTION Risk of injury caused by lamp rupture! Lamp rupture (audible as a loud bang) may lead to jamming of the lamp module and/or failure of the electronics modules.
  • Page 206 What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Switching to the backup lamp • Turn off the suspension system at the power switch before switching to the backup lamp. The lamp module contains two xenon lamps. The second lamp is used as a backup lamp which can be swung into the illumination beam path when the first lamp fails.
  • Page 207 S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Version 6.2 G-30-1607-en Page 207...
  • Page 208: Failure Of Magnetic Brakes

    What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Failure of magnetic brakes If the magnetic brakes fail (magnetic brakes are locked), you can manually po- sition the articulated arm including the microscope by overcoming the locking effect of the magnetic brakes.
  • Page 209: Failure Of The Zoom Function

    S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Failure of the zoom function • Adjust the magnification manually using zoom knob (1). • If the zoom drive moves into an end position, press the "MANUAL" button on the control panel of your suspension system.
  • Page 210: Failure Of The Focusing Function

    What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Failure of the focusing function • Adjust the focus manually using focusing knob (1). • If the focus drive moves into an end position, press the "MANUAL" button on the control panel of your suspension system.
  • Page 211: Causes Of Malfunctions And Remedies

    S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Causes of malfunctions and remedies For your safety • This system is a high-grade technological product. To ensure optimum performance and safe working order, we recommend having it checked by our service representative as part of regular scheduled maintenance.
  • Page 212 What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Problem Possible cause Remedy Zoom system inoperative. Defective motor. Adjust magnification manu- ally using the zoom knob. Contact service. Zoom system always moves Switch suspension system to to upper or lower end posi- MANUAL“...
  • Page 213 S88, S8, S81 / OPMI Vario What to do in the event of malfunctions SpeedFokus (Autofocus) option Fault Possible cause Remedy Cross-ref- erence SpeedFokus (Autofocus) op- Camera control unit (CCU) Switch on MediLive controller tion is inoperative switched off. (CCU). Error code: E20 Monitor switched off Switch on the monitor.
  • Page 214 What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Fault Possible cause Remedy Cross-ref- erence Focusing cannot be activated Different function assigned to Activate configuration mode Page 191 handgrip button 1 or 2 1 and assign the AF (Auto- focus) function to handgrip button 1 or 2.
  • Page 215 S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Suspension system Problem Possible cause Remedy No function at all. Line power failure. Contact in-house electrician. Power switch of suspension Press power switch. system not switched on. Automatic circuit breaker in Press power switch.
  • Page 216 What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Xenon illumination system Problem Possible cause Remedy Surgical field illumination on Thermal cut-out activated. Remove the cause of over- microscope not working. heating. For example, drapes could be covering the grid.
  • Page 217 S88, S8, S81 / OPMI Vario What to do in the event of malfunctions Problem Possible cause Remedy Xenon lamp is lit, but beep Defective lamp control Switch off the illumination. Page 181 sounds intermittently. system. Illuminate surgical field using an OR illuminator.
  • Page 218 What to do in the event of malfunctions S88, S8, S81 / OPMI Vario Video monitor Problem Possible cause Remedy No image display. Main switch is switched off. Switch on the main switch. Power cable not correctly Plug in the power cable. plugged in.
  • Page 219 S88, S8, S81 / OPMI Vario Care and Maintenance Care and Maintenance Care of the device................220 Cleaning ....................220 Sterilization .....................221 Disinfection .....................221 Changing the xenon lamp module............222 Balancing the monitor arm ..............224 Spare parts ....................226 Technical safety check ................227 Disposal....................228 Version 6.2 G-30-1607-en Page 219...
  • Page 220: Care Of The Device

    For the regular cleaning of objective lenses and eyepieces of the surgical mi- croscope, we recommend the optics cleaning set available from ZEISS. For the catalog number, please see the section "System data - Ordering data". Prevention of fogging To protect the eyepiece optics from fogging, we recommend using an anti- fogging agent.
  • Page 221: Sterilization

    S88, S8, S81 / OPMI Vario Care and Maintenance Sterilization The asepsis sets available from Carl Zeiss contain rubber caps and hand grips which can be sterilized in autoclaves. For detailed information on sterilization please see the enclosed instructions "Preparation of resterilizable products"...
  • Page 222: Changing The Xenon Lamp Module

    • Pack the old lamp module (5) in the transport packaging of the new lamp module. Fill in the enclosed return card and send the old lamp module to the nearest Zeiss service representative. • Only use the original transport case (2), as it also provides explosion pro- tection should Xenon lamps be defective.
  • Page 223 S88, S8, S81 / OPMI Vario Care and Maintenance Version 6.2 G-30-1607-en Page 223...
  • Page 224: Balancing The Monitor Arm

    Care and Maintenance S88, S8, S81 / OPMI Vario Balancing the monitor arm If the video monitor does not remain in place in every position required, the following readjustments can be performed. Increasing the friction of left/right movement of the carrier arm. •...
  • Page 225 S88, S8, S81 / OPMI Vario Care and Maintenance Version 6.2 G-30-1607-en Page 225...
  • Page 226: Spare Parts

    Care and Maintenance S88, S8, S81 / OPMI Vario Spare parts Surgical microscope Description Cat. no. Light guide, 2.5 m long, with 90° adapter 303481-9225-000 Video Description Cat. no. Y/C connection cable, 2 m long (2x) 301687-9102-000 Y/C extension cable, 12.5 m long 301687-9110-000 BNC (VBS) connection cable, 10 m long (2x) 301687-9101-000...
  • Page 227: Technical Safety Check

    S88, S8, S81 / OPMI Vario Care and Maintenance Technical safety check CAUTION Risk of injury! • Make sure that the regular technical safety checks required for this sys- tem in accordance with the applicable national regulations are performed on schedule and to the stipulated extent. To prevent any impairment of the system's safety as a result of ageing, wear, etc., the organization operating the system must ensure, in accordance with the applicable national regulations, that the regular technical safety checks...
  • Page 228: Disposal

    Care and Maintenance S88, S8, S81 / OPMI Vario Disposal User information on the disposal of electrical and electronic devices This symbol means that the product must not be disposed of as normal do- mestic waste. The correct disposal of electrical or electronic devices helps to protect the en- vironment and to prevent potential hazards to the environment and/or human health which may occur as a result of improper handling of the devices con- cerned.
  • Page 229 S88, S8, S81 / OPMI Vario System Data System Data Technical data ................231 Surgical microscope.................231 Suspension systems .................232 Ordering data ..................247 Regulatory information ..............251 Ambient conditions ............... 254 EMC (electromagnetic compatibility) ..........255 Electromagnetic interference ..............256 Electromagnetic immunity for ME equipment and ME systems ....257 Electromagnetic immunity for non-life-supporting ME equipment and ME systems..................259 Recommended safety distances between portable and mobile...
  • Page 230 System Data S88, S8, S81 / OPMI Vario Version 6.2 Page 230 G-30-1607-en...
  • Page 231: Technical Data

    S88, S8, S81 / OPMI Vario System Data Technical data Surgical microscope Component Features Magnification Motorized zoom system, zoom ratio 1:6, magnification factor γ = 0.4x - 2.4x. Total magnification 2.4x - 14.6x at a working distance of 200 mm with 10x eyepiece: 1.5x - 8.7x at a working distance of 415 mm Total magnification...
  • Page 232: Suspension Systems

    System Data S88, S8, S81 / OPMI Vario Suspension systems S88 floor stand Mechanics Component Features Suspension arm Length: 850 mm Swivel angle: 320° Vertical lift: ±360 mm Carrier arm Length: 450 mm Swivel angle: 320° Coupling Swivel angle: 450° Stand height 1880 mm Base...
  • Page 233 S88, S8, S81 / OPMI Vario System Data Version 6.2 G-30-1607-en Page 233...
  • Page 234 System Data S88, S8, S81 / OPMI Vario S88 floor stand with lifting column Mechanics Component Features Suspension arm Length: 850 mm Swivel angle: 320° Vertical lift: ±360 mm Carrier arm Length: 450 mm Swivel angle: 320° Coupling Swivel angle: 450° Stand height 1880 mm ...
  • Page 235 S88, S8, S81 / OPMI Vario System Data Version 6.2 G-30-1607-en Page 235...
  • Page 236 System Data S88, S8, S81 / OPMI Vario Electrical data of S88 floor stand Power connection Only connect the suspension system to wall outlets which are provided with a properly connected pro- tective ground conductor. Rated voltage 115 V∼ (100 - 120 V~) 230 V~ (220 - 240 V~) Current consumption 115 V~ max.1000 VA 230 V~ max.
  • Page 237 S88, S8, S81 / OPMI Vario System Data Xenon illumination system Fiber optic illumina- Xenon short-arc reflector lamp tion Color temperature: approx. 5000 K Rated power: approx. 180 W Backup lamp in lamp housing, manually selectable. Video monitor LCD display TFT color monitor Screen size 15"...
  • Page 238 System Data S88, S8, S81 / OPMI Vario DVI mode,VGA mode, Component mode Resolution Refresh rate (HZ) Description 640 x 480 640 x 480 720 x 576 720 x 576 800 x 600 800 x 600 VESA 800 x 600 VESA 800 x 600 VESA...
  • Page 239 S88, S8, S81 / OPMI Vario System Data Carrier arm for video monitor Mechanics Component Features Suspension arm length: 354mm swivel angle: 360° vertical lift: ±220mm Carrier arm length: 180 mm swivel angle: ±90° Admissible max. load 9.0kg on the monitor arm Total weight Approx.
  • Page 240 System Data S88, S8, S81 / OPMI Vario S8 ceiling mount Mechanics Component Features Suspension arm Length: 930 mm Swivel angle: 270° Vertical lift: ±360 mm Lift arm Length: 1000 mm Swivel angle: 350° Carrier arm Swivel angle of control panel: 180° (±90°) Coupling Swivel angle: 450°...
  • Page 241 S88, S8, S81 / OPMI Vario System Data Version 6.2 G-30-1607-en Page 241...
  • Page 242 System Data S88, S8, S81 / OPMI Vario Electrical data of S8 ceiling stand Rated voltage 115 V~ (100 - 120 V~) 230 V~ (220 - 240 V~) Current consumption 115 V: max. 750 VA 230 V: max. 750 VA Rated frequency 50...60 Hz Fuses...
  • Page 243 S88, S8, S81 / OPMI Vario System Data Version 6.2 G-30-1607-en Page 243...
  • Page 244 System Data S88, S8, S81 / OPMI Vario S81 ceiling mount Mechanics Component Features Suspension arm Length: 930 mm Swivel angle: 270° Vertical lift: ±360 mm Carrier arm Length: 600 mm Swivel angle: 350° Swivel angle of control panel: 180° (±90°) Coupling Swivel angle: 450°...
  • Page 245 S88, S8, S81 / OPMI Vario System Data Version 6.2 G-30-1607-en Page 245...
  • Page 246 System Data S88, S8, S81 / OPMI Vario Electrical data of S81 ceiling mount Rated voltage 115 V~ (100 - 120 V~) 230 V~ (220 - 240 V~) Current consumption 230 V~ max. 750 VA 120 V~ max. 750 VA Rated frequency 50...60 Hz Fuses...
  • Page 247: Ordering Data

    System Data Ordering data Only operate the system with the accessories included in the delivery package and approved by Carl Zeiss. You will find the contact responsible for orders in your country on this website: www.meditec.zeiss.com The product configurations, accessories and spare parts that are applicable for authorized Carl Zeiss partners are stipulated by contract.
  • Page 248 System Data S88, S8, S81 / OPMI Vario Suspension systems Description Cat. no. S88 floor stand 000000-1154-525 S88 floor stand with lifting column 000000-1169-820 S8 ceiling mount 000000-1176-968 S81 ceiling mount 000000-1176-969 Tubes Description Cat. no. Tiltable binocular tube, f = 170 mm, 180° 303791-0000-000 Tiltable binocular tube, f = 200 mm, 180°...
  • Page 249 S88, S8, S81 / OPMI Vario System Data Foot control panels Description Order number Wireless foot control panel, 14 functions 304970-9060-000 (FCP WL) Wired foot control panel, 14 functions 304970-9055-000 (FCP) Cable for foot control panel (FCP and FCP WL), 304970-8730-000 3 m cable length (recommended for floor stand) Cable for foot control panel (FCP and FCP WL),...
  • Page 250 Asepsis Description Cat. no. VisionGuard replacement lenses (sterile, 20/box) 306001-0000-000 Vision Guard lens protector (non-sterile, 10/box) 306002-0000-000 Zeiss sterile drapes, type 70, 306070-0000-000 pack of 5 Zeiss sterile drapes, type 71, 306071-0000-000 pack of 5 Zeiss sterile drapes, type 88,...
  • Page 251: Regulatory Information

    S88, S8, S81 / OPMI Vario System Data Regulatory information S88 / OPMI Vario Electrical safety The medical device complies with the requirements according to – IEC 60601-1:2005, – IEC 601-1:1988+A1:1991+A2:1995, – CAN/CSA-C22.2 No. 601.1-M90. Classification according to level of protection against electrical shock: Protection class 1 Protection rating (acc.
  • Page 252 System Data S88, S8, S81 / OPMI Vario S8 / OPMI Vario Electrical safety The medical device complies with the requirements according to – IEC 60601-1:2005, – IEC 601-1:1988+A1:1991+A2:1995, – CAN/CSA-C22.2 No. 601.1-M90. Classification according to level of protection against electrical shock: Protection class 1 Protection rating (acc.
  • Page 253 S88, S8, S81 / OPMI Vario System Data S81 / OPMI Vario Electrical safety The medical device complies with the requirements according to – IEC 60601-1:2005, – IEC 601-1:1988+A1:1991+A2:1995, – CAN/CSA-C22.2 No. 601.1-M90. Classification according to level of protection against electrical shock: Protection class 1 Protection rating (acc.
  • Page 254: Ambient Conditions

    System Data S88, S8, S81 / OPMI Vario Ambient conditions For operation Feature Permissible range Temperature + 10 °C ... + 40 °C Rel. Humidity 30 % ... 75 % Air pressure 700 hPa ... 1060 hPa For transportation and storage Feature Permissible range Temperature...
  • Page 255: Emc (Electromagnetic Compatibility)

    When using the device, the EMC precautions specified below must be ob- served. • Only use spare parts approved by Carl Zeiss for this device. • Do not use any portable or mobile RF communication equipment in the vicinity of the device as this may impair the device's function.
  • Page 256: Electromagnetic Interference

    System Data S88, S8, S81 / OPMI Vario Electromagnetic interference The S88, S8, S81 / OPMI Vario is intended for operation in an electromagnetic environment as specified below. The customer or the user of the device S88, S8, S81 / OPMI Vario is responsible for ensuring that the device is operated in such an environment.
  • Page 257: Electromagnetic Immunity For Me Equipment And Me Systems

    S88, S8, S81 / OPMI Vario System Data Electromagnetic immunity for ME equipment and ME systems The S88, S8, S81 / OPMI Vario is intended for operation in an electromagnetic environment as specified below. The customer or the user of the S88, S8, S81 / OPMI Vario is responsible for ensuring that the device is operated in such an environment.
  • Page 258 System Data S88, S8, S81 / OPMI Vario Immunity IEC 60601 - Compliance level Electromagnetic environment - tests test level guidelines Power frequency 3 A/m 3 A/m Power frequency magnetic fields (50/60Hz) mag- should be at levels typical of busi- netic field as per ness and hospital environments.
  • Page 259: Electromagnetic Immunity For Non-Life-Supporting Me Equipment And Me Systems

    S88, S8, S81 / OPMI Vario System Data Electromagnetic immunity for non-life-supporting ME equipment and ME systems The S88, S8, S81 / OPMI Vario is intended for operation in an electromagnetic environment as specified below. The customer or the user of the S88, S8, S81 / OPMI Vario is responsible for ensuring that the device is operated in such an environment.
  • Page 260 System Data S88, S8, S81 / OPMI Vario Immunity IEC 60601 - Compliance Electromagnetic environment - tests test level level guidelines Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is influ- enced by absorption and reflection by structures, objects and persons. Field strengths of stationary transmitters such as base stations for mobile telephones and mobile land radio equipment, amateur radio stations, AM and FM radio broadcast and TV broadcast transmitters cannot be theoretically predicted accurately.
  • Page 261: Recommended Safety Distances Between Portable And Mobile Rf Communication Equipment And The S88, S8, S81 / Opmi Vario

    S88, S8, S81 / OPMI Vario System Data Recommended safety distances between portable and mobile RF communication equipment and the S88, S8, S81 / OPMI Vario The S88, S8, S81 / OPMI Vario is intended for use in an electromagnetic envi- ronment in which RF disturbances are controlled.
  • Page 262: System Combinations

    ME system. Connecting electrical devices not approved by Carl Zeiss can lead to a reduced level of safety on the ME system. When assem- bling your ME system (Use of the power output socket), make sure to observe the requirements of IEC 60601-1:2005, Section 16, and note the following: •...
  • Page 263 S88, S8, S81 / OPMI Vario Index Index Numerics 180° tiltable tube ..................80 Address of manufacturer ................2 Adjusting the balance setting of the suspension arm ......152 Adjusting the eyepieces ................162 Adjusting the friction of the microscope's axis of rotation ......160 Adjusting the iinterpupillary distance ............162 Adjustment range, lamp brightness ............159 Adjustment range, motor speed .............159...
  • Page 264 Index S88, S8, S81 / OPMI Vario Button B ....................197 Carrier arm for the surgical microscope .............70 Carrier arm, back ..................92 CCD array, entering the size of ..............201 CCD position ...................189 Checklist ....................170 Chip size, entry of ...................201 Collision and crushing, risk of ..............102 Component connector (3x cinch) ............108 Composite video connector (1x cinch) .............108 Configuration mode 1 for OPMI Vario ............190...
  • Page 265 S88, S8, S81 / OPMI Vario Index Electrical standard ..................236 EMC (electromagnetic compatibility) ............255 Emergency backup power supply .............10 Eyecups ....................82 Eyepiece magnification, entry for calculating total magnification ....196 Eyepiece mount ..................80 Eyepiece optics, fog-free .................140 Eyepieces, mounting of ................138 Eyepieces, technical data ................231 Face-to-face coobservation ..............136 Failure of a function ................170...
  • Page 266 Index S88, S8, S81 / OPMI Vario Hazard symbols ..................7 Identical modules of the suspension systems ..........90 Illuminated field ..................74 Illuminated-field diameter ...............231 Illuminated-field knob ................74 Indicator window for rated voltage ..........100, 118 Information symbols ...................7 Instrument tray (option) ................102 Instrument tray, strap provided ...............102 Intended use ....................8 Key row ....................92...
  • Page 267 S88, S8, S81 / OPMI Vario Index Manual activation of the backup lamp ............84 Manual function ..................26 MANUAL key ..................92, 178 Manual mode ................25, 26, 178 Manufacturer .....................2 Manufacturer’s address ................2 Maximum configuration .................136 Microscope body ..................70 Minimum configuration ................136 Mobile phones ..................255 MODE key ....................178 Motor speed, adjustment range .............159...
  • Page 268 Index S88, S8, S81 / OPMI Vario Rating label ...................36, 42, 48 Recentering ....................78 Regulatory information ................251 Release bar ....................90 Release button for magnetic brakes ............76 Release buttons, freely programmable ............76 Relocating the unit .................148 Remedy ....................211 Remote socket ................100, 118, 130 Repeat function ..................178 Risk of burn injuries caused by high illumination intensity ......17 ROI display on/off ...................189...
  • Page 269 S88, S8, S81 / OPMI Vario Index Surgical microscope, adjustment ............162 Surgical microscope, connecting .............140 Suspension systems ................232 S-Video connector ..................108 Switching to the backup lamp ............24, 206 Symbols and labels on the device .............28 System overview ..................57 Target group ....................8 TFT monitor ....................106 Total magnification or working distance, OPMI Vario ......186 Trademarks ....................2...
  • Page 270 Index S88, S8, S81 / OPMI Vario X-Y recentering ..................190 XY und Z movement ................197 Yellow indicator lamp .................24, 88 Zoom function, failure ................209 Zoom knob ....................74 Zoom memory ..................190 Zoom release button .................76 Zoom speed, OPMI Vario ................187 Version 6.2 Page 270 G-30-1607-en...
  • Page 271 S88, S8, S81 / OPMI Vario (Blank page, for your notes . . .) Version 6.2 G-30-1607-en Page 271...
  • Page 272 Carl Zeiss Meditec AG Fax: + 49 (0)7364 - 20 4823 Goeschwitzer Strasse 51-52 Email: surgical@meditec.zeiss.com 07745 Jena Internet: www.meditec.zeiss.com Germany...

This manual is also suitable for:

Opmi vario s8Opmi vario s81

Table of Contents