Instructions For Use (Qualified Personnel) - Otto Bock AxonEnergy Integral 757B501 Instructions For Use Manual

Hide thumbs Also See for AxonEnergy Integral 757B501:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11
CAUTION
Remaining, operating and charging in areas outside the allowable temperature range
Injury due to faulty control or malfunction of the product.
Avoid remaining, operating and charging in areas outside the allowable temperature range
(see page 16).
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the product.
CAUTION
Independent manipulation of the product
Injury due to malfunction and resulting unexpected prosthesis actions.
Manipulations to the product other than the tasks described in these instructions for use are
not permitted.
The product and any damaged components may only be opened and repaired by authorised,
qualified Ottobock personnel.
CAUTION
Charging the product with soiled or damaged contacts
Injury due to unexpected behaviour of the product caused by insufficient charging.
Make sure that the contacts are always clean and free of grease.
Clean the electrical contacts of the charging plug and charging receptacle regularly using
cotton swabs and a mild soap solution.
Take care to avoid damaging the contacts with pointed or sharp objects.
CAUTION
Insufficient distance to HF communication devices (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference with internal data
communication.
Therefore, keeping a minimum distance of 30 cm from HF communication devices is recom­
mended.
5 Scope of delivery
1 pc. 757B501 AxonEnergy Integral
2 pc. lamination dummy
1 pc. lamination dummy – charging receptacle
1 pc. drilling template for charging receptacle
1 pc. Component holder set
1 pc. Instructions for use (qualified personnel)