Beep signal
1 x long
2 x short
3 x short
10.2.3 Warnings/error signals
Charging recept­
acle
LED is flashing red
LED flashes 3x in the
colour of the current
charge level
LED alternately
flashes yellow/red
10.3 Axon
The term "Axon" stands for Adaptive eXchange Of Neuroplacement data. The Axon-Bus is an
Ottobock innovation for the field of exoprosthetics: a data transmission system, derived from
safety-related bus systems in the aviation and automobile industries. For the user this means
enhanced safety and reliability because of a considerably reduced sensitivity to electromagnetic
interference in comparison with conventional systems.
1 Avant-propos
INFORMATION
Date de la dernière mise à jour : 2020-12-14
Veuillez lire attentivement l'intégralité de ce document avant d'utiliser le produit ainsi que
respecter les consignes de sécurité.
Apprenez à l'utilisateur comment utiliser son produit en toute sécurité.
18
Additional display
-
LED on charging
receptacle lights up
briefly
-
Beeper
3x short
-
3x short
Event
Turn off prosthesis on charging receptacle
Charging begins (charging plug connected
to charging receptacle)
Charging ends (charging plug disconnected
from charging receptacle)
Switch on prosthesis
Battery voltage too low, prosthesis shuts off auto­
matically
Explanation/rectification
Error in the Axon-Bus prosthesis solution
Turn the Axon-Bus prosthesis solution on/off
Check the plug connection in the socket
Connect the charger.
If the service icon (yellow wrench) or system
error icon (red icon) is displayed on the bat­
tery charger, send the Axon-Bus prosthesis
solution to an Ottobock Service Center.
The connection from the product to the Axon­
Master is interrupted
Check the connection (plug connection, flex
cable, Easy Plus, etc.)
Overheating of the Axon-Bus prosthesis
solution
Allow the Axon-Bus prosthesis solution to
cool down for a few minutes
Do not use the Axon-Bus prosthesis solution
at high ambient temperatures (Technical
data)