Wear protective clothing when removing or replacing the battery (disposable gloves and
protective goggles at a minimum).
Avoid direct skin contact!
CAUTION
Using the product with pointed or sharp objects
Destruction or damage to the product during installation.
Do not use any pointed or sharp objects when installing the product.
CAUTION
Damage to the product due to improper installation
Injury due to product damage.
Install the battery according to the steps in the "Laminating instructions" section (see fig. 5).
NOTICE
Improper care of the housing
Damage to the casing through the use of acetone, white spirit or similar solvents.
Only clean the housing with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).
NOTICE
Use of incorrect power supply unit/battery charger
Damage to product due to incorrect voltage, current or polarity.
Use only power supply units/battery chargers approved for this product by Ottobock (see
instructions for use and catalogues).
INFORMATION
Prior to initial operation of the product, charge the supplied battery.
4.4 Patient information
WARNING
Operating the product near active implanted systems
Interference with active implantable systems (e.g., pacemaker, defibrillator, etc.) due to electro­
magnetic interference of the product.
When operating the product in the immediate vicinity of active implantable systems, ensure
that the minimum distances stipulated by the manufacturer of the implant are observed.
Make sure to observe any operating conditions and safety instructions stipulated by the man­
ufacturer of the implant.
CAUTION
Mechanical stress on the product
Injury due to faulty control or malfunction of the product components.
Do not subject the product to mechanical vibrations or impacts.
Check the product for visible damage before each use.