Warnings/Error Signals - Otto Bock Genium X3 Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for Genium X3:
Table of Contents

Advertisement

Beep signal
1x short
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app or using the remote control, there are
no beep signals (see Page 31).
Beep signal
Vibration
signal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short

15.2.2 Warnings/error signals

Error during use
Beep signal
Vibration signal
1x before beep signal Self-test completed successfully, product is
Additional action per­
formed
Mode switching using the
Cockpit app or remote con­
trol
Bouncing on the forefoot fol­
lowed by unloading the
prosthetic leg
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Vibration signal
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen­
erated)
3 x long
5 x long
Event
operational
Event
Mode switching performed using
the Cockpit app or remote con­
trol.
Bouncing pattern recognised.
Switching to basic mode
(mode 1) carried out.
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydraulic
unit
Battery charge level
under 25%
Battery charge level
under 10%
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Ottobock | 43