Beep signal
5 x long
10 x long
30 x long
Error while charging the product
LED on
Status
power
LED on
supply
battery
charger
44 | Ottobock
Vibration signal
5 x long repeated
every 60 seconds
10 x long
1x long, 1x short
repeated every
3 seconds
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
Event
Medium error:
e.g. a sensor is not
operational
Switch to safety
mode (see Page 38).
Battery charge level
5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error
e.g. failure of valve
drives
Possibly no switching
into safety mode (see
Page 38).
Resolution
Check whether the country-specif­
ic plug adapter is fully engaged on
the power supply.
Check socket with another electric
appliance.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Required action
Restricted walking pos­
sible. The change in flex­
ion resistance has to be
noted.
The product must be
inspected immediately by
a prosthetist.
Charge the battery.
Restricted walking pos­
sible. The possible
change in flexion/exten­
sion resistance must be
observed.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, it is
no longer permissible to
use the product. The
product must be inspec­
ted immediately by a
prosthetist.

Advertisement