Appendices
Beep signal
Vibration sig­
nal
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short

17.2.2 Warnings/error signals

Error during use
Beep signal
5 x long
10 x long
30 x long
52
Additional action performed
Bouncing on the forefoot followed
by unloading the prosthetic leg
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Vibration signal
1 x long at intervals of
approx.
5 seconds
mute mode (silent mode)
is activated, this signal is
not generated)
3 x long
5 x long
5 x long repeated every
60 seconds
10 x long
1 x long, 1 x short
repeated every 3 seconds
Continuous
Event
Bouncing pattern recognised.
Switching to basic mode (mode 1) car­
ried out.
Switching to MyMode 1 (mode 2) car­
ried out.
Switching to MyMode 2 (mode 3) car­
ried out.
Result
Overheated hydraulic unit Reduce activity.
(if
Charge level under 25%
Charge level under 10%
Medium
error
Page 45)
e.g. a sensor is not oper­
ational
Battery charge level 5%
After the beep and vibra­
tion signals, the product
switches to empty battery
mode and then switches
off.
Severe error/indication
of safety mode activa­
tion (see Page 45)
e.g. one or more sensors
are not operational.
Total failure
Electronic
control
longer possible. Safety
mode active or undeter­
mined
valve
Undetermined
product
behaviour.
Required action
Charge battery soon.
Remaining
operating
approx. 24 hours
Charge battery soon
Remaining
operating
approx. 6 hours
(see
Walking possible with restric­
tions. Please note the change
in flexion resistance.
The product must be inspec­
ted immediately by an author­
ised Ottobock Service Center.
Charge the battery.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product is prohibited. The
product must be inspected by
an authorised Ottobock Ser­
vice Center.
Attempt to reset this error by
connecting/disconnecting the
no
battery charger.
If the error persists, use of the
product is prohibited. The
state.
product must be inspected by
an authorised Ottobock Ser­
vice Center.
Genium X3 3B5-3/3B5-3=ST