Beep signal
Vibration
signal
1x short
1x short
Short at peri
Short at
odic intervals
periodic
intervals
1x long
1x long
14.2.2 Warnings/error signals
Error during use
Beep signal
–
–
–
10x long
30x long
88
Additional action per
formed
User has sat on the bicycle
ergometer and commenced
pedaling movement
The pedaling movements
were continued.
The prosthetic leg was
extended or the foot was
placed on the floor.
Vibration signal
1x long at interval of
approx. 5 seconds
3x long
5x long
10x long
1x long, 1x short
repeated every
3 seconds
Event
After a few pedaling movements,
this was recognised and switch
ing to the "2.Bicycle ergometer"
MyMode took place.
The flexion and extension resist
ances are reduced to the extent
that the knee joint moves freely.
Placing the foot on the floor was
recognised and switching back to
the "1. Basic Mode" MyMode
took place.
Event
Hydraulics over
heated
Charge level under
25%
Charge level under
15%
Charge level 0%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic
ation of safety
mode activation
For example, sensor
not ready for opera
tion or valve drive fail
ure
Possibly no switching
into safety mode.
Required action
Reduce activity.
Charge battery soon.
Charge battery immedi
ately; the product will be
switched off after the next
warning signal.
Charge the battery.
Walking possible with
restrictions. Please note
the possible change in
flexion/extension resist
ance.
Attempt to reset this error
by connecting/discon
necting the battery char
ger. The battery charger
must remain connected
for at least 5 seconds
before it is disconnected.
If the error persists, use
of the product is prohib
ited. The product must be
inspected by an O&P
professional immediately.