Beep signal
Vibration
signal
2x short
2x short
3x short
3x short
16.2.2 Warnings/error signals
Error during use
Beep signal
–
–
–
5x long
10x long
30x long
Additional action per
formed
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Vibration signal
1x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
3x long
5x long
5x long repeated
every 60 seconds
10x long
1x long, 1x short
repeated every
3 seconds
Event
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydrau
lics
Battery charge level
under 25%
Battery charge level
under 10%
Error of moderate
severity (see
page 87)
e.g. a sensor is not
operational
Charge level 5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic
ation of safety
mode activation
(see page 87)
e.g. one or more
sensors are not oper
ational
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Note the
change in flexion resist
ance.
The product must be
inspected by an O&P
professional immediately.
Charge the battery.
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, use
of the product is prohib
ited. The product must be
inspected by an O&P
professional immediately.
95