Warnings/Error Signals - Otto Bock Genium X3 3B5-3 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-3:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 53
Beep signal
Vibration
signal
2x short
2x short
3x short
3x short

16.2.2 Warnings/error signals

Error during use
Beep signal
5x long
10x long
30x long
Additional action per­
formed
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Vibration signal
1x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen­
erated)
3x long
5x long
5x long repeated
every 60 seconds
10x long
1x long, 1x short
repeated every
3 seconds
Event
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydrau­
lics
Battery charge level
under 25%
Battery charge level
under 10%
Error of moderate
severity (see
page 87)
e.g. a sensor is not
operational
Charge level 5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
(see page 87)
e.g. one or more
sensors are not oper­
ational
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Note the
change in flexion resist­
ance.
The product must be
inspected by an O&P
professional immediately.
Charge the battery.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected by an O&P
professional immediately.