Warnings/Error Signals - Otto Bock Genium 3B1 Instructions For Use Manual

Bionic prosthetic system
Hide thumbs Also See for Genium 3B1:
Table of Contents

Advertisement

Available languages

Available languages

16.2.2 Warnings/error signals

Error during use
Beep signal
5 x long
10 x long
30 x long
100
Vibration signal
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen­
erated)
3 x long
5 x long
5 x long repeated
every 60 seconds
10 x long
1 x long, 1 x short
repeated every
3 seconds
Continuous
Event
Overheated hydraulic
unit
Charge level under
25%
Charge level under
10%
Medium error (see
Page 92)
e.g. a sensor is not
operational
Battery charge level
5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
(see Page 92)
e.g. one or more
sensors are not oper­
ational
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Please note
the change in flexion res­
istance.
The product must be
inspected immediately by
an O&P professional.
Charge the battery.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected immediately by
an O&P professional.

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Genium 3b1-stGenium 3b1-2Genium 3b1-2-st

Table of Contents