Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see
page 30).
Beep signal
Vibration
signal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short
16.2.2 Warnings/error signals
Error during use
Beep signal
–
–
–
5x long
46
Additional action per
formed
Mode switching using the
Cockpit app
Bouncing on the forefoot fol
lowed by unloading the
prosthetic leg
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Vibration signal
1x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
3x long
5x long
5x long repeated
every 60 seconds
Event
Mode switching is performed
using the Cockpit app.
Bouncing pattern recognised.
Switching to basic mode
(mode 1) carried out.
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydrau
lics
Battery charge level
under 25%
Battery charge level
under 10%
Error of moderate
severity (see
page 39)
e.g. a sensor is not
operational
Required action
Reduce activity.
Charge battery soon.
Remaining operating time
approx. 24 hours
Charge battery soon
Remaining operating time
approx. 6 hours
Walking possible with
restrictions. Note the
change in flexion resist
ance.
The product must be
inspected by an O&P
professional immediately.