Beep signal
Vibration sig
nal
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short
16.2.2 Warnings/error signals
Error during use
Beep signal
5x long
10x long
30x long
Genium 3B1/3B1=ST
Additional action performed
Bouncing on the forefoot followed
by unloading the prosthetic leg
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Weight taken off prosthetic leg
and leg kept still for 1 second
Vibration signal
1x long at intervals of
approx.
5 seconds
mute mode (silent mode)
is activated, this signal is
not generated)
3x long
5x long
10x long
1x long, 1x short repeated
every 3 seconds
Event
Bouncing pattern recognised.
Switching to basic mode (mode 1) car
ried out.
Switching to MyMode 1 (mode 2) car
ried out.
Switching to MyMode 2 (mode 3) car
ried out.
Event
Overheated hydraulic unit Reduce activity.
(if
Battery
charge
under 25%
Battery
charge
under 10%
Error
of
moderate
severity:
e.g. a sensor is not oper
ational
Switch to safety mode
(see Page 86).
Battery charge level 5%
After the beep and vibra
tion signal, the product
switches to empty battery
mode and then switches
off.
Severe error
e.g. failure of valve drives
Possibly no switching into
safety
mode
Page 86).
Required action
level
Charge battery soon.
Remaining
operating
approx. 24 hours
level
Charge battery soon
Remaining
operating
approx. 6 hours
Restricted walking possible.
The change in flexion resist
ance has to be noted.
The product must be inspec
ted immediately by an author
ised Ottobock Service Centre.
Charge the battery.
Restricted walking possible.
The possible change in flex
ion/extension resistance must
be noted.
(see
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, it is no
longer permissible to use the
product. The product must be
inspected by an authorised
Ottobock Service Centre.
Appendices
time
time
Ottobock | 91