17.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration sig
nals.
17.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
—
1x short
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see
Page 83).
Beep signal
Vibration
signal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short
17.2.2 Warnings/error signals
Error during use
Beep signal
–
Vibration signal
3x long
1x short
Additional action per
formed
Mode switching using the
Cockpit app
Bouncing on the forefoot fol
lowed by unloading the
prosthetic leg
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Vibration signal
1 x long at intervals of
approx. 5 seconds (if
mute mode (silent
mode) is activated,
this signal is not gen
erated)
Event
Charging mode started (3 sec. after connecting
the battery charger)
Self-test completed successfully, product is
operational
Event
Mode switching is performed
using the Cockpit app.
Bouncing pattern recognised.
Switching to basic mode
(mode 1) carried out.
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
Event
Overheated hydraulic
unit
Required action
Reduce activity.
99