Warnings/Error Signals - Otto Bock Genium Instructions For Use Manual

Hide thumbs Also See for Genium:
Table of Contents

Advertisement

Available languages

Available languages

Appendices

17.2.2 Warnings/error signals

Error during use
Beep signal
5 x long
10 x long
30 x long
Error while charging the product
LED on
Status LED
power sup­
on battery
ply
charger
100
Vibration signal
1 x long at intervals of
approx.
5 seconds
mute mode (silent mode)
is activated, this signal is
not generated)
3 x long
5 x long
5 x long repeated every
60 seconds
10 x long
1 x long, 1 x short
repeated every 3 seconds
Continuous
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
Result
Overheated hydraulic unit Reduce activity.
(if
Charge level under 25%
Charge level under 10%
Medium
error
Page 93)
e.g. a sensor is not oper­
ational
Battery charge level 5%
After the beep and vibra­
tion signals, the product
switches to empty battery
mode and then switches
off.
Severe error/indication
of safety mode activa­
tion (see Page 93)
e.g. one or more sensors
are not operational.
Total failure
Electronic
control
longer possible. Safety
mode active or undeter­
mined
valve
Undetermined
product
behaviour.
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check socket with another electric appli­
ance.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Required action
Charge battery soon.
Remaining
operating
approx. 24 hours
Charge battery soon
Remaining
operating
approx. 6 hours
(see
Walking possible with restric­
tions. Please note the change
in flexion resistance.
The product must be inspec­
ted immediately by an author­
ised Ottobock Service Center.
Charge the battery.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product is prohibited. The
product must be inspected by
an authorised Ottobock Ser­
vice Center.
Attempt to reset this error by
connecting/disconnecting the
no
battery charger.
If the error persists, use of the
product is prohibited. The
state.
product must be inspected by
an authorised Ottobock Ser­
vice Center.
Genium 3B1-3/3B1-3=ST