Calibrating the temperature probe

102 / 198
1.
In the [Basic setup] menu select the menu item  Language.
2.
Select your language using the arrow keys.
3.
Press the Enter key to confirm your selection.
In the [Temp. unit] menu you can select in which unit the temperature value
is displayed. In general, this setting is valid for all windows in the display.
1.
In the [Basic setup] menu select the menu item  Temp. unit.
2.
Select one of the following options:
n
n
3.
Press the Enter key to confirm your selection.
If, when checking the temperature in a steady state, you discover a constant
temperature deviation of T
can be equaled out via the Calibration menu point.
With the menu point Offset (1-point comparison), the characteristic of the
temperature switch is adjusted in parallel by the input value.
With the menu point 2-point calibration (2-point comparison), the charac-
teristic of the temperature switch is adjusted and the slope of the character-
istic is changed.
Integral Process Thermostats and High-Temperature Thermostats
A list containing the languages opens.
The new setting is valid with immediate effect.
A list of the options opens.
With [°C] all temperatures are displayed in °Celsius .
With [°F] all temperatures are displayed in °Fahrenheit .
The new setting is valid with immediate effect.
A calibrated reference thermometer with the desired level of accu-
racy is necessary. Otherwise you should not change the calibration
of your constant temperature equipment.
or T
from the reference thermometer, this
int
ext
It is possible to change the temperature values T
range of ±3 K respectively.
and T
within a
int
ext