Setting the volume of the acoustic signals

The LAUDA Proline Thermostats signal alarms as a dual-tone acoustic signal and warnings as a
continuous tone.
Command
Alarm
Pump
Menu
7.6.6

Entering the offset of the internal temperature probe

If, during checking with a calibrated reference thermometer (e.g. from the LAUDA DigiCal Series) a
deviation is found, then the offset (i.e. the additive part of the characteristic) of the internal measuring
chain can be adjusted with the following function. The reference thermometer must be dipped into the
bath according to the details on the calibration certificate.
Command
intern Pt100
extern Pt100
Pump
Menu
YAUE0007 / 10/2/2018
loud
medium
low
off
End
T
T
set
Calibration
Default
End
T
T
set
Starting up
Sounds
 Open the device parameter menu via the
soft key
 With the cursor keys continue to
Settings
Sounds
 Select either
 Example on left:
 Select the desired volume with
.
 Accept selection with
fix
quit the window with
changes.
Calibration
 Open the device parameter menu via the
soft key
 With the cursor keys continue to
Settings
intern Pt100
 The window shown on the left appears.
 Confirm selection with
fix
Menu .
Basic Settings
.
Alarm
or
Warning
Alarm
is set to loud.
or End or
without making
Menu .
Calibration
Calibration
.
.
.
or
65