Lauda IN 130 T Operation Manual page 151

Integral process thermostats and high-temperature thermostats
Table of Contents

Advertisement

Fault
Display: Gage pressure
(Outflow pressure too high)
Display: Evaporator frozen
Unstable temperature control
(Temperature fluctuates strongly)
V6
Integral Process Thermostats and High-Temperature Thermostats
Cause ð Possible remedy
Pump level is too high ð select a lower pump level.
n
When pressure control is active, the set pressure is too
n
high ð reduce the set pressure.
The maximum pressure is too low ð increase the max-
n
imum pressure, but only if your application, hoses, etc.
cannot be damaged.
Increase the pump level.
n
Open the manual bypass in the constant temperature
n
equipment.
Increase the flow rate through the application.
n
Increase the pump level.
n
Open the manual bypass in the constant temperature
n
equipment.
If necessary, increase the control parameters Xp and Tn.
n
Increase the flow rate through the application.
n
Reduce the load.
n
151 / 198