Lauda IN 130 T Operation Manual page 56

Integral process thermostats and high-temperature thermostats
Table of Contents

Advertisement

ID Function
57 Control parameter XpF
61 Control parameter Prop_E
Table 17: Control
ID Function
59 Set point offset
67 Control in control variable X: 0 = internal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt 2 (only for Integral)
69 Offset source X for set point: 0 = normal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt 2 (only for Integral)
Table 18: Rights
ID Function
63 Status of Master keyboard: 0 = free / 1 = blocked
65 Status of remote control keyboard: 0 = free / 1 = blocked
Table 19: Status
ID Function
75 Status of standby: 0 = Device is switched on / 1 = Device is switched off
107 Device type (e.g.: "INT" or "VC")
130 Device status: 0 = OK / -1 = fault
131 Fault diagnosis; a 7-digit answer in the format XXXXXXX is output, whereby
each character X contains fault information (0 = no fault / 1 = fault).
The following information is defined for the seven places of the answer
format:
1st character = Error
n
2nd character = Alarm
n
3rd character = Warning
n
4th character = Overtemperature
n
5th character = Low level
n
6th character = 0 (with alarm setting: High level
n
7th character = External control value missing
n
56 / 198
Integral Process Thermostats and High-Temperature Thermostats
Unit
Command
[–]
IN_PAR_10
[K]
IN_PAR_15
Unit
Command
[K]
IN_PAR_14
[–]
IN_MODE_01
[–]
IN_MODE_04
Unit
Command
[–]
IN_MODE_00
[–]
IN_MODE_03
Unit
Command
[–]
IN_MODE_02
TYPE
[–]
STATUS
[–]
STAT
V6

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents