ID Function
52 Control parameter TdE
54 Correction limitation
56 Control parameter XpF
60 Control parameter Prop_E
Table 28: Control
ID Function
58 Set point offset
66 Control in control variable X: 0 = internal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2 (only for Integral)
68 Offset source X for set point: 0 = normal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2
Note (ID 66 and 68): If X = 3, the commands ID 66 and ID 68 cannot be executed in some constant temperature control
devices until an external temperature specification has been received (via the command ID 15).
Table 29: Rights
ID Function
62 Master keyboard (equivalent to "KEY"): 0 = unlock / 1 = lock
64 Keyboard remote control unit (command): 0 = unlock / 1 = lock
Table 30: Status
ID Function
74 Switch the device on / off (standby)
Table 31: Programmer
ID Function
76 Select the program for the subsequent commands (X = 1 – 5). The
default program is 5 when the constant temperature equipment is
switched on.
Caution!
Executing this command may stop any programs that are currently
running.
78 Start programmer
79 Pause programmer
60 / 198
Integral Process Thermostats and High-Temperature Thermostats
Unit
Command
[s]
OUT_PAR_07_XXXX.X
[K]
OUT_PAR_09_XXX.X
[–]
OUT_PAR_10_XX.X
[K]
OUT_PAR_15_XXX
Unit
Command
[K]
OUT_PAR_14_XXX.X
[–]
OUT_MODE_01_X
[–]
OUT_MODE_04_X
Unit
Command
[–]
OUT_MODE_00_X
[–]
OUT_MODE_03_X
Unit
Command
[–]
START / STOP
Unit
Command
[–]
RMP_SELECT_X
[–]
RMP_START
[–]
RMP_PAUSE