Lauda IN 130 T Operation Manual page 49

Integral process thermostats and high-temperature thermostats
Table of Contents

Advertisement

Fig. 21: Example for entering the ping command
Check the LAN network and the process
interface
V6
2.
There are two ways of checking:
n
n
It is easy to check the connection to the interface with a PC and Microsoft
Windows operating system.
For Windows 3.11 with the program "Terminal".
n
For Windows 95/98/NT/XP the program "HyperTerminal".
n
For operating systems Windows Vista, Windows 7, Windows 8 and
n
Windows 10 "HyperTerminal"* is not part of the operating system.
Integral Process Thermostats and High-Temperature Thermostats
Enter the ping command together with the IP address.
ping XXX.XXX.XXX.XXX@
For "XXX.XXX.XXX.XXX", put the IP address that was entered
when the Ethernet interface was configured.
Or
Enter the ping command together with the serial number of the
thermostatic circulator (possible with software control system
1.36 or later).
ping serial number@
If the Ethernet interface is configured and connected correctly,
the interface will return four responses within a very short time.
See Fig. 21.
* Terminal programs are available on the Internet as freeware. These
programs offer features similar to "HyperTerminal" (for example
PuTTY or RealTerm). Search query "serial port terminal program".
49 / 198

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents