Lauda IN 130 T Operation Manual page 162

Integral process thermostats and high-temperature thermostats
Table of Contents

Advertisement

Installation location
Integral T
Table 1
ACC area À
Temperature stability Á
Device dimensions
width x depth x height
Filling volume
‑ minimum
‑ maximum
Pump data 50 Hz
‑ maximum discharge pres-
sure
‑ Maximum flow rate
Pump data 60 Hz
‑ maximum discharge pres-
sure
‑ Maximum flow rate
Connector thread (outer)
inlet/outlet
Noise level 50 Hz Â
Noise level 60 Hz Â
Weight
Distance between device and
environment
‑ Front
‑ Back
162 / 198
Ensure that adequate ventilation is provided
n
Minimum free room volume, room volume per kg of refrigerant
n
according to DIN 378-1
(For refrigerant quantity, see type plate or Ä Chapter 15.4 "Refrigerant
and filling weight" on page 174).
Table 48: Refrigerant and room volume
Refrigerant
R-449A
R-452A
R-23
R-508B
Unit
IN 130 T
°C
-30 – 120
K
±0.05
mm
430 x 550 x 760 430 x 550 x 760 430 x 550 x 760
L
3.6
L
8.7
bar
3.5
l/min
40
bar
4.0
l/min
45
Inches
G¾"
dB(A)
61
dB(A)
61
kg
79.0
mm
500
mm
500
Integral Process Thermostats and High-Temperature Thermostats
Room volume per kg of refrigerant
2.81 m³/kg
2.37 m³/kg
1.48 m³/kg
4.0 m³/kg
IN 230 T
IN 230 TW
-30 – 120
-30 – 120
±0.05
±0.05
3.6
3.6
8.7
8.7
3.5
3.5
40
40
4.0
4.0
45
45
G¾"
G¾"
63
58
63
60
84.0
85.0
500
200
500
200
IN 530 T
-30 – 120
±0.05
560 x 550 x
1325
7.2
20.5
3.5
40.0
4.6
45
G¾"
62
66
148.5
500
500
V6

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents