Fig. 61: Program 1
Available settings
Fig. 62: In program editor
V6
3.
The following options are available:
n
n
n
4.
Select the  Edit menu item.
You can stop the programmer by pressing the [STOP] softkey. After
pressing the [START]softkey, the programmer continues to run in the previ-
ously selected mode (pause or active mode).
Setting
No.
Tend
hh
:mm
Tolerance
Integral Process Thermostats and High-Temperature Thermostats
[Status]
Select the option [Start] to start the program.
Once the program has started, you can press [Hold] to pause
it.
A paused program can be continued by pressing [Continue].
Select the option [Stop] to end the program.
[Edit]
[Cycles]
Here you enter the number of repetitions of the selected
program.
The program appears on the display and you can now edit it.
100 temperature-time programs can be saved. A maximum of 250
freely programmable segments can be used per program.
The programmer encoder can be controlled or modified using the
timer.
Description
Program segment number
End temperature that should be reached
Time in hours (hh) by which the specified tem-
perature should be reached
Time in minutes (: mm) in which the specified tem-
perature should be reached
Tolerance defines how close the temperature
should be to the set temperature before the next
segment is processed.
0.0 means that there is no tolerance. This means
that the program starts the next temperature after
the preset time, even if the output temperature has
not yet been reached.
109 / 198