Table 13: Actuating signal
ID Function
11 Resolution of controller actuating signal in per mill
– negative value à device is cooling
– positive value à device is heating
13 Controller actuating signal in watts
– negative value à device is cooling
– positive value à device is heating
Table 14: Cooling
ID Function
24 Cooling mode: 0 = off / 1 = on / 2 = automatic
Table 15: Safety
ID Function
35 Timeout communication via interface (1 – 60 seconds; 0 = Off)
73 Status of Safe Mode: 0 = off (inactive) / 1 = on (active)
162 Heating output restriction
166 Heating output restriction in Safe Mode
168 Pump pressure set point in Safe Mode
170 Flow rate set point in Safe Mode, only for device with flow controller
172 Pressure limitation in Safe Mode, only for device with flow controller
Table 16: Control parameters
ID Function
39 Control parameter Xp
41 Control parameter Tn (181 = Off)
43 Control parameter Tv
45 Control parameter Td
47 Control parameter KpE
49 Control parameter TnE
51 Control parameter TvE
53 Control parameter TdE
55 Correction limitation
V6
Integral Process Thermostats and High-Temperature Thermostats
Unit, resolu-
Command
tion
IN_PV_06
[‰]
[W]
IN_PV_08
Unit
Command
[–]
IN_SP_02
Unit
Command
[s]
IN_SP_08
[–]
IN_MODE_06
[%]
IN_SP_13
[%]
IN_SP_14
[bar]
IN_SP_15
[l/min]
IN_SP_16
[bar]
IN_SP_17
Unit
Command
[–]
IN_PAR_00
[s]
IN_PAR_01
[s]
IN_PAR_02
[s]
IN_PAR_03
[–]
IN_PAR_04
[s]
IN_PAR_05
[s]
IN_PAR_06
[s]
IN_PAR_07
[K]
IN_PAR_09
55 / 198