ID Function
80 Continue programmer (after pause)
81 End programmer

4.7.8 Error messages

V6
The error messages of the Ethernet interface are described in the following.
The string ERR_X or ERR_XX is output after an incorrect command.
Error
ERR_2
ERR_3
ERR_5
ERR_6
ERR_8
ERR_30
ERR_31
ERR_32
ERR_33
ERR_34
ERR_35
ERR_36
ERR_37
ERR_38
ERR_39
ERR_40
ERR_41
Integral Process Thermostats and High-Temperature Thermostats
Unit
[–]
[–]
Description
Incorrect entry (for example, buffer overflow)
Wrong command
Syntax error in value
Impermissible value
Module or value not available
Programmer, all segments occupied
Not possible to specify a set point value. The "Setpoint offset"
function is active.
TiH £ TiL
External sensor missing
Analog value not present
Automatically configured
Not possible to specify set point, programmer is running or
has been paused
Not possible to start programmer (analog set point value input
is ON)
The operator does not have the privileges to execute the
command. Another control section has exclusive operating
rights, which means that writing via this interface is not per-
mitted.
Operation is not permitted. Safe Mode is active.
Operation is not permitted. Safe Mode is switched off.
Operation is not permitted. Constant temperature equipment
is in error status.
Command
RMP_CONT
RMP_STOP