7.6.5 Setting the volume of the acoustic signals
The LAUDA Proline Thermostats signal alarms as a dual-tone acoustic signal and warnings as a continuous tone.
Command
Alarm
Pump
Menu
7.6.6 Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer a deviation is found, then the offset (i.e. the additive part
of the characteristic) of the internal measuring chain can be adjusted with the following function. The reference
thermometer must be dipped into the bath according to the details on the calibration certificate.
Command
intern Pt100
extern Pt100
Pump
Menu
62 / 137
loud
medium
low
off
End
T
T
set
Calibration
Default
End
T
T
set
Proline Kryomats
Sounds
Open the device parameter menu via the soft
key
Menu .
With the cursor keys continue to
Settings
Select either
Example on left:
Select the desired volume with
.
Accept selection with
fix
window with
Calibration
Open the device parameter menu via the soft
key
Menu .
With the cursor keys continue to
Settings
intern Pt100
The window shown on the left appears.
Confirm selection with
fix
Basic Settings
Sounds
Alarm
or
Warning
.
Alarm
is set to loud.
or
or End or quit the
without making changes.
Calibration
Calibration
.
.
03/2022
.