682 Synergy Plus
6.3

EDP Command Calibration

Use the following instructions to calibrate the 682 using EDP commands. For information on the EDP commands of the 682,
see
Section 12.0 on page
84. Access to Setup mode
NOTE: The indicator must respond with OK after each step or the calibration procedure must be done again.
For commands ending with #s, s is the scale number (1).
1. Put the indicator in Setup mode
2. For a standard calibration, remove all weight from scale (except hooks or chains which are needed to attach weights).
3. Send the command SC.WZERO#s to perform a standard calibration of the zero point.
• Send SC.TEMPZERO#s to perform a temporary zero calibration
• Send SC.LASTZERO#s to perform a last zero calibration
4. Apply the span calibration weight to the scale.
5. Send the command SC.WVAL#s=xxxxx, where xxxxx is the value of the span calibration weight applied to the scale.
6. Send the command SC.WSPAN#s to calibrate the span point. Continue on to
points, or proceed to
7. Apply weight equal to the first linear point to the scale.
8. Send the command SC.WLIN.Vn#s=xxxxx, where n is the linear point number (1-4) and xxxxx is the exact value of
the weight applied.
9. Send the command SC.WLIN.Cn#s to calibrate the linear point, where n is the linear point number (1-4).
10. Repeat
steps 7–9
for up to four total linear points.
11. If hooks or chains were used to attach the weights, remove all weight, including the hooks and chains, and send the
command SC.REZERO#s to remove the zero offset.
12. Send the command KSAVEEXIT to return to Weigh mode.
58
Visit our website
(Section 4.1 on page
(Section 4.1 on page
step
11.
www.RiceLake.com
29) is required.
29).
step 7
to calibrate additional linear

Advertisement