Edp Command Calibration - Rice Lake Synergy 680HE Technical Manual

Hostile environment digital weight indicator
Table of Contents

Advertisement

680HE Synergy Indicator
5.3

EDP Command Calibration

Use the following instructions to calibrate the 680HE using EDP commands. For information on the EDP commands of the
680HE, see
Section 7.0 on page
The indicator must respond with OK after each step or the calibration procedure must be done again.
Note
1. Place indicator into setup mode via the setup jumper
2. For a standard calibration, remove all weight from scale (except hooks or chains which are needed to attach weights).
3. Send the command SC.WZERO#s to perform a standard calibration of the zero point.
• Send SC.TEMPZERO#s to perform a temporary zero calibration
• Send SC.LASTZERO#s to perform a last zero calibration
4. Apply the span calibration weight to the scale.
5. Send the command SC.WVAL#s=xxxxx, where xxxxx is the exact value of the span calibration weight applied to the
scale.
6. Send the command SC.WSPAN#s to calibrate the span point. Continue on to
linearization points, or proceed to
7. Apply weight equal to the first linearization point to the scale.
8. Send the command SC.WLINVn#s=xxxxx, where n is the linearization point number (1-4) and xxxxx is the exact
value of the weight applied.
9. Send the command SC.WLIN.Cn#s to calibrate the linearization point.
10. Repeat
steps 7–9
for up to four total linearization points.
11. If hooks or chains were used to attach the weights, remove all weight, including the hooks and chains, and send the
command SC.REZERO#s to remove the zero offset.
12. Send the command KSAVEEXIT to return to weigh mode.
42
Visit our website
44.
step
11.
www.RiceLake.com
(Section 4.1 on page
21).