6-2 Detailed assembly procedures

· Before assembling the microscope, set the main switch of the control box STM7-CB/STM7-CBA to " " (OFF)
CAUTION
and unplug the power cord.
· The frame and the stage are very heavy. Pay a careful attention when carrying them. For the number of transport
persons, refer to page 5.
} Unless otherwise specified, the same procedures apply to STM7-SF/STM7-SFA/STM7-MF/STM7-MFA/STM7-LF/STM7-LFA.
b
1
Frame installation
CAUTION
} If there is no sufficient space around the table, you can use a
movable desk instead.
a
1
Take out the sleeve a and the protecting sheet b from the package
box.
2
Hold the transport sticks b (4 pcs.) to raise the frame and place it
on the installation table.
(Frame weight: 51.8 kg (STM7-SF), 53.8 kg (STM7-SFA), 77.1 kg (STM7-
MF), 78.6 kg (STM7-MFA), 150.5 kg (STM7-LF), 152 kg (STM7-LFA))
CAUTION
} If the installation table is dented by the legs of the frame due to
the thin top board, or if the microscope is influenced by vibration
significantly, place the anti vibration plates STM7-VI (option) under
c
the legs of the frame.
} Be sure to place the legs of the frame in the center of the anti
vibration plates.
· Keep the transport sticks and clamps/screws for Z-axis
transport lock in a safe place. They will be used for next
transportation.
· Place the frame on a stable horizontal surface.
· Secure the space (1 m) around the table for workers. If
there is no sufficient space, the frame cannot be installed.
· Do not hold any part other than the transport sticks. The
microscope frame may be damaged.
STM7