82
8
9
10
11
12
6
Move the X-axis to right and left ends.
3
6
7
Repeat from
to
until the edge of the measurement object always
comes to the center of the cross line.
8
Rotate the eyepiece sleeve along the edge of the objective.
9
Tighten the clamping knob with the tool provided with the frame to fix
the eyepiece with cross hairs.
10
Slide the light path selector knob to move to the pulled out position.
The light path is 100% TV/Photo.
11
While observing the live image on the monitor, rotate the camera so
that the cross line fits the edge.
12
Fix the camera adapter by tightening the clamping screw with the
Allen screwdriver provided with the frame.