Page 1
SZX10 Research High-Class Stereo Microscope This instruction manual is for the Olympus SZX10 Research High-Class Stereo Microscope System. Optical Microscope and Accessory To ensure the safety, obtain optimum performance and to familiarize yourself full with the use of this microscope, we recommend that you study this manual thoroughly before operating the microscope.
Page 2
Downloaded from ManualsNet.com search engine...
SZX10 Important Safety Precautions 1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection. · Moving this microscope is accompanied with the risk of dropping the specimen and causing infections.
Page 6
Getting Ready 1. A microscope is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact. 2. Do not use the microscope where it is subjected to direct sunlight, high temperature and humidity, or vibration. (For the operating environment, see “6 Specifications”...
Page 7
SZX10 Maintenance and Storage 1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze). 2. If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.
This enables the large base to be used as is. However, be sure to use the drop prevention collar (SZX-R) in combination. Olympus classifies SZX2-ZB10 as an optical microscope and other units as optical microscope accessories. Downloaded from ManualsNet.com...
SZX10 Controls If the microscope is not yet assembled, see “8 Assembly” (P. 19) before the following. Light path selector knob (P.13) Trinocular tube only. Zooming knob Diopter adjustment ring (P.10) 0.63X to 6.3X ±5 m (per meter) Coarse focus adjustment knob rotation tension adjustment ring (P.7)
Summary Of Observation Procedure 3-1 Preparation Ref. Page 1. Check and tighten the connection of each component, especially the observation tube................(Page 21 ) 2. Check that the angled formed by the microscope body with respect to the base is less than the turning-over prevention angle............................................(Page 20 ) 3.
SZX10 Using The Controls 4-1 Base Using the Stage Plate In reflected light observation, the stage plate can be placed either the white or black side facing up. Win transmitted light observation, use the transparent NOTE glass stage plate (SP-C).
Engaging and Disengaging the Zooming Knob Click Stop Function When the click stop knob is set to ON, the click stop function is engaged for each magnification indicated with the zooming knob. When the knob is set to OFF, the zoom magnification can be varied continuously and finely near the click groove.
SZX10 Zoom Magnification Indication The body zoom magnification a is indicated on the zooming knob on the right side. The total magnification of observation can be calculated with the following formula: Objective Body zoom Eyepiece magnification magnification magnification The magnification indication ring of an objective other NOTE than the 1X objective cannot be used.
4-3 Observation Tube Adjusting the Interpupillary Distance Be sure to hold the binocular assembly a with both NOTE hands to make this adjustment. To prevent damaging the mechanism, do not apply an excessive force beyond the stop position. While looking through the eyepieces, hold the left and right of the binocular assembly a and adjust the eyepieces by opening or closing them for binocular vision until the left and right fields of view coincide completely.
SZX10 When Using the Eyepiece Micrometer Disk Look into the right eyepiece that includes the eyepiece micrometer disk, and focus the micrometer disk by rotating the diopter adjustment ring b . Place an easy-to-observe specimen on the stage plate. Rotate the zooming knob c to the highest magnification and, looking only into the right eyepiece, focus the specimen u s i n g t h e c o a r s e a n d f i n e f o c u s a d j u s t m e n t k n o b s .
Mounting the Eyepiece Micrometer Disk A variety of eyepiece micrometer disks a can be inserted into the WHSZ10X-H, WHSZ15X-H and WHSZ20X-H eyepieces. Please purchase an eyepiece micrometer disk with a diameter of 24 mm and thickness of 1.5 mm. Unscrew the micrometer disk frame b by turning it counterclockwise to remove it from the bottom of the eyepiece.
SZX10 Selecting the Light Path of the Trinocular Tube Pull out the light path selector knob a to set the light path for the right eyepiece to the camera light path. (The following table shows the intensity ratio of each light path in %.)
4-4 Camera Observation and Photomicrography The camera observation and photomicrography are available when the SZX2-TR30/TR30PT/TTR/TTRPT trinocular observation tube is used. A digital camera unit can be mounted on the straight tube mount of the trinocular tube by means of the camera adapter and/or camera mount adapter*.
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, contact Olympus. Problem...
Allen wrench** (provided with the standard base and focusing assembly) Objective Objective series for SZX10 * The Allen screwdriver can be stored in the pillar. Allen screwdriver storage hole* ** Do not use the Allen wrench for a purpose other than installing the pillar and the zoom Stage plate microscope body.
8-2 Detailed Assembly Procedure Installing the Pillar Using the Allen wrench provided with the base, loosen the two clamping screws a on the pillar support sleeve completely. Hold the pillar b so that the Allen screwdriver storage hole made of white rubber comes at the top, and insert the pillar into the pillar support sleeve until it reaches the bottom.
Page 25
SZX10 Mounting the Microscope Body Remove the cap a on the focusing assembly by inserting a thin object into the notch. Using the Allen wrench, loosen the dovetail mount clamping screw inside the cap on the focusing assembly by rotating it by 2 or 3 turns (counterclockwise).
Page 26
Mounting the Eyepiece When using the micrometer eyepiece disk, insert it in the right eyepiece. (It may be inserted in the left eyepiece but the description in this manual assumes that it is inserted in the right eyepiece.) Remove the eyepiece dust caps a and loosen the eyepiece clamping screws b completely.
SZX10 Operation Of Other Modules 9-1 Eyepoint Adjuster SZX-EPA This module increases the eye point height by 40 mm. Up to two eyepoint adjusters can be mounted, provided that no other intermediate attachment is used. External View Observation tube clamping screw...
9-2 Drop Prevention Collar SZX-R and Auxiliary Pillar SZH-P400/SZH-P600 · The auxiliary pillar is to be used when observing a large specimen or when it is required to move the microscope body upward for using a low-magnification objective with long WD. ·...
SZX10 9-3 Revolving Nosepiece SZX2-2RE10 The revolving nosepiece allows two objectives to be mounted. This makes objective switching easy by simply rotating the revolving nosepiece and expands the range of observation magnifications. External View Hand guard Allen wrench (for M3 screws)
Page 30
Place the revolving nosepiece mount c (with its objective mounts d facing upward) where the objective mount has been by aligning the screw holes. Using the Allen wrench (for M3 screws), clamp the revolver mount with the provided three clamping screws (M3, 6 mm long) e . As the screw holes may be hidden behind the objective NOTE mounts d , clamp the nosepiece mount while rotating it.
Page 31
SZX10 Return the microscope body to the illumination light path by loosening the focusing module clamping knob b , rotating the microscope body gently in the direction of the arrow, aligning visually the objective with the illumination light path and tightening the clamping knob b again.
9-4 BX Stage Adapter Type 1 SZX-STAD1 This adapter is for installation of a U-SRG or U-SRP rotary stage on the SZX2 standard base or a SZX2 series illumination base. When the U-SRP rotary stage is used together with the U-FMP mechanical stage, X-Y directional movement becomes possible, which is convenient for framing during photomicrography.
SZX10 9-5 BX Stage Adapter Type 2 SZX-STAD2 This adapter is for installing the U-SIC4R2/SIC4L2 large stage* on the SZX2-STL large stand**. When this stage adapter is used, the SZH-P400 auxiliary pillar should be used to cover the height of the stage adapter (about 125 mm).
Page 34
Simplified Transmitted Light Observation Illuminate the specimen with an external light source (LSD illuminator, Microscope body light guide illuminator, etc.). Objective Light the external light source as shown in the figure on the left and Frosted filter irradiate the mirror assembly. Stage Eliminate irregularities in illumination.
SZX10 9-6 Stage Adapter Type 1 SZH-STAD1 This adapter has the same function as the SZX-STAD1 BX stage adapter type 1, but the usable stages with this adapter are the BH2-SH horizontal knob stage. Assembly STAD mount for ILLT SZX2-STADM is necessary for using the SZX2-ILLTQ/ILLTS LED transmitted light illumination base.
9-7 Gliding stage SZH-SG External view and nomenclature Stage plate Finger hook Seat Gliding stage Gliding stage Illumination field diameter: 40 mm Moving range diameter: 40 mm Installation · Be sure to clean the sliding surface if dirt or metallic power is attached on them. NOTE ·...
SZX10 9-8 Cup stage SZH-SC External view and nomenclature The SZH-SC can be used only with reflected light illumination. It cannot be used with transmitted light NOTE illumination. Tube Cup stage Finger hook Tilting limit angle: 30° Stage plate Specimen holder...
Page 38
Operation Place a specimen on the stage plate, hold the cup stage by the edge, and tilt the cup stage slowly. If the specimen slips on the stage plate, hold the specimen with the provided specimen holder. To fix a container such as a petri dish, insert the provided tube into the specimen holder to fix the container by pinching.
Page 39
Downloaded from ManualsNet.com search engine...
Page 40
Manufactured by Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan Distributed by AX7480 11 Issued in March, 2019 Downloaded from ManualsNet.com search engine...