12. Do not operate the clutch lever of the stage while it is moving (page
36). Otherwise, accuracy deterioration or damage may result.
13. To maintain accuracy, it is recommended that the microscope
undergoes the Olympus periodic inspection service (accuracy
inspection, calibration, functional check, etc.) once a year. It is
recommended that the microscope also undergoes this service
when the microscope has been reinstalled at a different location.
(Note that this service is subject to charge.)
14. In case of emergency, unplug the power cord to stop supplying
electricity.
Install the microscope at a location where you can reach the power
cord connector (rear of the control box) or the outlet at hand to
disconnect the power cord.
15. The basic software (STM7-BSW) can also be used with the
measuring microscope STM6 (with some restrictions).
16. The hardware of STM6 is not compatible with the hardware of STM7.
If modules other than modules described in this instruction manual are
connected to the microscope, the system may be damaged.
Emergency stop switch
1. The emergency stop switch b on the control box for motorized
STM7-CBA forces the Z-axis movement to stop. (motorized frame
only)
In case of emergency, press the emergency stop switch.
2. To release the switch, rotate the emergency stop switch in the
resetting direction (clockwise), and then rotate the fine/coarse
focusing knob of the focus controller STM7-MCZ to move the
focusing unit to the safe position.
Then, set the main switch of the control box for motorized to " "
(OFF), and set the main switch of the control box for motorized to "
" (ON) again.
When the main switch is set to " " (ON), five short beeps indicating
an error are generated. (When using the digital indicator STM7-DI,
the display shows "013F1507". To cancel this error, operate either the
FOCUS buttons
or the fine/coarse focusing knob of the focus
controller to move the Z-axis by 50 μm or more and set the main
switch to " " (OFF) and then to " " (ON).
3. If the emergency stop switch is pressed during the origin search,
you cannot move the focusing unit even though the emergency
stop switch is rotated to the resetting direction.
To cancel this, while the emergency stop switch is rotated to
the resetting direction, set the main switch of the control box for
motorized to " " (OFF) and then set the main switch to " " (ON)
again. When the error sound (five short beeps) is heard (If you are
using the digital indicator STM7-DI, "013F1507" is displayed on the
display area.), cancel the error according to the procedures in 2.
Height Measurement
1. To measure the height with higher accuracy, use an objective with
high magnification. (50x or higher)
STM7
7