Problem
7. Stage
a) The observation image moves
too much when the stage is
touched with hand.
b) The X- and/or Y-feed knobs
run idle.
c) The stage moves
spontaneously when a clutch
is released.
d) The specimen is moved when
a knob is turned quickly.
8. Display (Digital indicator, Basic software and Excel transfer tool)
a) The display shows nothing.
b) The Excel transfer tool does
not start.
c) The measured values are
too different from what is
expected.
d) The values cannot be
transferred to the controller.
e) Only the X- and Y-axis values
are output to the external
processing devices.
f) Display shows "013F0500".
g) Display shows "013F1503".
h) Display shows "013F1504".
i) Display shows "013F1505".
j) Display shows "013F1506".
k) Display shows "013F1507".
92
Cause
A stage clutch lever is released.
The modules (stage, etc.) are loosened.
The microscope is used in an
environment with frequent vibration.
The microscope is installed on a table
with a thin top board.
A clutch lever is released.
A clutch is set to an intermediate position. Release the clutches and retry setting
The horizontality of the base is not
adjusted.
The specimen is moved at the stage
movement speed
The main switch of the control box is not
set to ON.
The fuses in the control box are blown.
Cables are not connected completely.
Cables are not connected completely.
The COM port is used for other functions.
The "μm/mm/inch/mil" setting is
erroneous.
Microsoft Office is not started.
The version other than Microsoft Office
2007, 2010, 2013, 2016, or 2019 is installed.
The data output is set to the X-/Y-axes
only.
The setting value is not inputted.
The hand switch STM7-HS is not
connected to the control box for manual
STM7-CB.
The focus controller STM7-MCZ is not
connected to the control box for motorized
STM7-CBA.
The Z-scale cable is disconnected.
The emergency stop switch is pressed.
Since the emergency stop switch was
pressed before exiting last time, the
confirmation of operations when the
microscope starts is interrupted.
Remedy
Set both clutches securely.
Secure the modules firmly.
Install the microscope in an
environment with less vibration.
Install the microscope on a table with
a thick top board.
Set the clutches securely.
them securely.
The horizontality of the base is not
adjusted.
Adjust the horizontality of the desk
where the frame is installed.
Fix the specimen.
Set the main switch to " " (ON).
Replace the fuses.
Connect them properly.
Connect them properly.
Check the settings of the COM port
again.
Set it properly.
Start Microsoft Office.
Install the version later than
Microsoft Office 2007.
Set it properly.
Input the setting value.
Connect the cable.
Connect the cable.
Connect the cable.
Reset the emergency stop switch.
Operate either the FOCUS button
of the focus controller, the AF
START/STOP button or the fine/coarse
focusing knob.
Page
36
36
36
62
22
73
69
69
49
49
74
69
69
69
22
22