6-3 Inputting setting values (maker setting mode)

· Setting values must be inputted by Olympus. Do not change the setting values by yourselves.
CAUTION
· Values of "Standard value" and "Measured value" described in Certificate of Accuracy are setting values
assured when shipping the product. If the setting values are not inputted or the values different from
those described in Certificate of Accuracy are set, the accuracy is not assured even in the accuracy
assurance environments. For accuracy assurance environments, refer to page 89.
} When the microscope is assembled, input the setting values. If the setting values are not inputted, an error sound
beeps (three short beeps) when the main switch of the control box STM7-CB/STM7-CBA is set to " "(ON). (If you are
using the control box for motorized STM7-CBA and also the origin search setting is ON, the error sound beeps
when the origin search is finished.)
} To input the setting values, the values of X-, Y- and Z-axis described in Certificate of Accuracy provided with the
frame or the stage are required. Prepare Certificate of Accuracy at hand.
· Do not change the items not related to settings. If you change the factory default settings, the system
CAUTION
may not count or the motorized frame may not be driven in Z direction, etc. However, this is not a
failure of the microscope.
If you change the settings by mistake, start the maker setting mode to set the correct settings, save
the setting values and exit.
} When you are not using the digital indicator STM7-DI, the maker setting mode can be started with the Excel transfer
tool downloaded from the Olympus website. For details, refer to the Help of the Excel transfer tool.
74