f
g
STM7-CS200/STM7-CS300
i
STM7-CS50/STM7-CS100
60
2
In case of STM7-CS200 or STM7-CS300, take out the inner box f
from the packaging box.
3
Take out the protecting sheet g from the packaging box.
4
In case of STM7-CS200 or STM7-CS300, take out the stage by holding
the transport sticks h (4 positions).
In case of STM7-CS50 or STM7-CS100, take out the stage by holding
the transport rods i (2 positions).
h
(Stage weight: 18.1 kg (STM7-CS50), 17.7 kg (STM7-CS100),
59.8 kg (STM7-CS200), 111.3 kg (STM7-CS300))
· Do not hold any part other than transport rods or transport
CAUTION
sticks. The stage may be damaged.
· Remove the anti-rust sheets on the top surface and the
bottom surface of the stage.
5
Place only the half of the stage on the installation table and wipe the
attaching surface clean.
In STM7-CS50 or STM7-CS100, the Y-axis knob is
CAUTION
protruded below from the stage mounting surface. Be
careful not to allow the Y-axis to touch the installation
table.
6
Place the attaching surface wiped on the stage attaching surface of
the frame. Wipe the other half of the attaching surface clean.
While wiping the stage, one or more installation workers
CAUTION
must support the stage by holding the transport rods or
the transport sticks.
7
Place the stage gently on the frame by aligning the mounting surfaces.
} While looking from the stage glass, fit the center of the stage with
the center of the window lens for transmitted light illumination of the
frame.
When placing the stage, be careful not to make the stage
CAUTION
collide with the glass surface of the frame.