Measurement in TRACK mode

f
j
h
1
i
2
3
4
5
6
Approximately focus on the specimen (page 28)
Press the AF MODE button a on the focus controller STM7-MCZ.
(LASER LED b on the front of the autofocus unit is turned ON and
SHOT of the digital indicator STM7-DI blinks to be in the AF standby
status.)
Turn ON the objective magnification display LED d on the front of
the autofocus unit you want to use by using the AF OB button c
of the focus controller. Then, engage the objective displayed into the
light path.
Press the AF MODE button a on the focus controller.
(The [AF TRACK] indicator e on the front of the control box for
motorized is turned ON and TRACK of the digital indicator blinks to
be in the TRACK mode.)
Bring the point at which you want to detect focusing to the center of
crosslines of the eyepiece (page 36).
Press the AF START/STOP button
(TRACK of the digital indicator is turned ON to execute AF in the
TRACK mode.)
When the focusing position is detected, short beeps indicating the beginning
of tracking are generated and tracking at the focusing position starts.
(FOCUS LED g on the front of the autofocus unit is turned ON.)
When the focusing position cannot be detected, three short beeps
indicating an error are generated and FOCUS LED g on the front
of the autofocus unit blinks.
When moving the focusing unit closer to the focusing position by
rotating the FOCUS button h of the focus controller or the coarse
focusing knob i , the focusing position is detected automatically
and it tracks at the focusing position after the one short beep
indicating the start of tracking is generated.